No response or low<br />
response to gas input<br />
Slow response to gas input<br />
Erratic reading<br />
Non-zero background<br />
reading<br />
“CAL Err” message during<br />
calibration<br />
Flashing “CAL” message<br />
during normal operation<br />
Readings consistently<br />
wrong when exposed to gas<br />
During normal operation,<br />
“pump” message flashing<br />
“xxx installed<br />
bias error” message on<br />
LCD<br />
Calibrates OK, but LEL or<br />
VOC readings off by<br />
constant percentage.<br />
“Lamp” message during<br />
operation<br />
1) sensor input is blocked or dirty filter<br />
2) defective sensor module<br />
3) defective electronics<br />
4) weak, or leaking pump<br />
5) sensor cover plate missing, loose, or misaligned<br />
6) empty sensor position<br />
1) sensor input is blocked or dirty filter<br />
2) defective sensor module<br />
3) sensor has inherently slow response time<br />
4) weak or leaking pump<br />
5) sensor cover plate missing, loose, or misaligned<br />
6) empty sensor position<br />
1) incorrect calibration<br />
2) defective sensor module<br />
3) defective electronics<br />
4) weak or leaking pump<br />
1) dirty filter<br />
2) incorrect zero calibration or zero drift<br />
3) defective sensor module<br />
4) non-zero background<br />
5) carbon filter saturated or not installed<br />
1) incorrect calibration<br />
2) low gas pressure in gas cylinder<br />
3) defective sensor module or weak lamp<br />
4) leak in flow path<br />
5) defective electronics<br />
1) Test 2.1<br />
2) Test 2.4<br />
3) Test 2.4<br />
4) Test 2.2<br />
5) Test 2.1<br />
6) Test 2.1<br />
1) Test 2.1<br />
2) Test 2.4<br />
3) Test 2.4<br />
4) Test 2.2<br />
5) Test 2.1<br />
6) Test 2.1<br />
1).Test 2.5<br />
2) Test 2.4<br />
3) Test 2.4<br />
4) Test 2.2<br />
1) Test 2.1<br />
2) Test 2.6<br />
3) Test 2.4<br />
4) Test 2.6<br />
5) Test 2.1<br />
1) Test 2.5<br />
2) Test 2.5<br />
3) Test 2.4<br />
4) Test 2.1<br />
5) Test 2.4<br />
1) excessive negative zero drift of sensor 1) Test 2.6<br />
1) incorrect span calibration<br />
2) defective sensor module<br />
3) incorrect correction factor<br />
1) Gas inlet blocked or no connection to inlet port<br />
2) weak or inoperative pump<br />
3) pump stall threshold needs adjustment<br />
1) bias switch in wrong position for Toxic1 sensor<br />
2) defective sensor module<br />
1) Wrong calibration or measurement gas selection<br />
2) Incorrect lamp selected<br />
3) incorrect dilution ratio<br />
1) dirty PID sensor or lamp<br />
2) weak or defective lamp<br />
3) Lamp fail threshold adjustment needed<br />
1) Test 2.5<br />
2) Test 2.4<br />
3) Test 4.5<br />
1) Test 2.1<br />
2) Test 2.2<br />
3) Test 2.7<br />
1) Test 2.8<br />
2) Test 2.4<br />
1) Test 4.5<br />
2) Test 4.5<br />
3) Test 4.5<br />
1) Test 2.9<br />
2) Test 2.9<br />
3) Test 2.9<br />
3) Electronics:<br />
Problem Possible Causes Test Procedures<br />
LED, back light, buzzer or 1) defective component 1) Test 3.2<br />
vibrator alarm inoperative<br />
Missing segment on the<br />
LCD<br />
1) defective LCD<br />
2) temperature too low<br />
3) contrast set too low<br />
1) Test 3.3<br />
2) Test 3.3<br />
3).Test 3.3<br />