No response or low<br />

response to gas input<br />

Slow response to gas input<br />

Erratic reading<br />

Non-zero background<br />

reading<br />

“CAL Err” message during<br />

calibration<br />

Flashing “CAL” message<br />

during normal operation<br />

Readings consistently<br />

wrong when exposed to gas<br />

During normal operation,<br />

“pump” message flashing<br />

“xxx installed<br />

bias error” message on<br />

LCD<br />

Calibrates OK, but LEL or<br />

VOC readings off by<br />

constant percentage.<br />

“Lamp” message during<br />

operation<br />

1) sensor input is blocked or dirty filter<br />

2) defective sensor module<br />

3) defective electronics<br />

4) weak, or leaking pump<br />

5) sensor cover plate missing, loose, or misaligned<br />

6) empty sensor position<br />

1) sensor input is blocked or dirty filter<br />

2) defective sensor module<br />

3) sensor has inherently slow response time<br />

4) weak or leaking pump<br />

5) sensor cover plate missing, loose, or misaligned<br />

6) empty sensor position<br />

1) incorrect calibration<br />

2) defective sensor module<br />

3) defective electronics<br />

4) weak or leaking pump<br />

1) dirty filter<br />

2) incorrect zero calibration or zero drift<br />

3) defective sensor module<br />

4) non-zero background<br />

5) carbon filter saturated or not installed<br />

1) incorrect calibration<br />

2) low gas pressure in gas cylinder<br />

3) defective sensor module or weak lamp<br />

4) leak in flow path<br />

5) defective electronics<br />

1) Test 2.1<br />

2) Test 2.4<br />

3) Test 2.4<br />

4) Test 2.2<br />

5) Test 2.1<br />

6) Test 2.1<br />

1) Test 2.1<br />

2) Test 2.4<br />

3) Test 2.4<br />

4) Test 2.2<br />

5) Test 2.1<br />

6) Test 2.1<br />

1).Test 2.5<br />

2) Test 2.4<br />

3) Test 2.4<br />

4) Test 2.2<br />

1) Test 2.1<br />

2) Test 2.6<br />

3) Test 2.4<br />

4) Test 2.6<br />

5) Test 2.1<br />

1) Test 2.5<br />

2) Test 2.5<br />

3) Test 2.4<br />

4) Test 2.1<br />

5) Test 2.4<br />

1) excessive negative zero drift of sensor 1) Test 2.6<br />

1) incorrect span calibration<br />

2) defective sensor module<br />

3) incorrect correction factor<br />

1) Gas inlet blocked or no connection to inlet port<br />

2) weak or inoperative pump<br />

3) pump stall threshold needs adjustment<br />

1) bias switch in wrong position for Toxic1 sensor<br />

2) defective sensor module<br />

1) Wrong calibration or measurement gas selection<br />

2) Incorrect lamp selected<br />

3) incorrect dilution ratio<br />

1) dirty PID sensor or lamp<br />

2) weak or defective lamp<br />

3) Lamp fail threshold adjustment needed<br />

1) Test 2.5<br />

2) Test 2.4<br />

3) Test 4.5<br />

1) Test 2.1<br />

2) Test 2.2<br />

3) Test 2.7<br />

1) Test 2.8<br />

2) Test 2.4<br />

1) Test 4.5<br />

2) Test 4.5<br />

3) Test 4.5<br />

1) Test 2.9<br />

2) Test 2.9<br />

3) Test 2.9<br />

3) Electronics:<br />

Problem Possible Causes Test Procedures<br />

LED, back light, buzzer or 1) defective component 1) Test 3.2<br />

vibrator alarm inoperative<br />

Missing segment on the<br />

LCD<br />

1) defective LCD<br />

2) temperature too low<br />

3) contrast set too low<br />

1) Test 3.3<br />

2) Test 3.3<br />

3).Test 3.3<br />

4