Whole body 3 Tesla Magnetic Resonance Imaging System optimized for higher performance in Cardiac, Body<br />

and neuroradiological examination with short super conducting magnet, High performance Gradient and<br />

Digital radi<strong>of</strong>requency. All capabilities are detailed below should be integral part <strong>of</strong> the quotation and none <strong>of</strong><br />

these essential requirements should be quoted as optional. If supplier has any advanced application or<br />

techniques available with them, the same may be quoted as optional.<br />

1 Magnet System<br />

a. Magnet should be superconductive with a field strength <strong>of</strong> 3T+ 0.12T<br />

b. It should have 70cm patient bore after positioning <strong>of</strong> gradient, shim, and RF antena. The magnet should have the<br />

facilities <strong>of</strong> better illumination, ventilation and designed to avoid patient claustrophobia<br />

c. Magnet should be shielded from external interference. 5 gauss line should within the confine <strong>of</strong> the magnet room<br />

d. Homogeneity <strong>of</strong> the magnet should be mentioned in relation to 10,20,30,40, and 45 cm DSV, Give details <strong>of</strong><br />

number <strong>of</strong> plane plots and number <strong>of</strong> measurements per plane to measure the homogeneity<br />

e. Global and local auto shimming should be available. Automated patient specific on line shimming should be<br />

available. Shim channels or any other method to correct the patient induced field distortions must be present.<br />

Facility <strong>of</strong> high order shimming should be available.<br />

f. Specify the weight <strong>of</strong> the magnet including gradients and cover etc..<br />

g. The front panel display at the magnet should display coil table position<br />

2 Gradient System<br />

1. Active shielded gradient system with strength <strong>of</strong> at least 40mT/m or more with the slew rate <strong>of</strong> 200mT/m/sec or<br />

more. This slew rate <strong>of</strong> 200mt/m/sec at 40 mT/m gradient should be available in each axis independent, for over all<br />

better duty cycle performance <strong>of</strong> the gradient.<br />

2. Duty cycle should be 100 percent, Give details.<br />

3. Gradient should have the provision for eddy current compensation.<br />

4. Gradient should be capable <strong>of</strong> delivering the desired clinical performance.<br />

a. Largest field <strong>of</strong> view should be at least 40 cm in all axis.<br />

b.Minimum TE in Gradient Echo 2D/3D should be specified for all sequences.<br />

c. Minimum slice thickness in 2D should be specified.<br />

d. Minimum slice thickness in 3D should be specified.<br />

e. Maximum Echo Train Length in both SE and GRE should be at least 256.<br />

f. Measurement matrix should be from 128X128 to 512x512 in both 2D and as well as 3D imaging.<br />

3 RF System<br />

1 RF system should be fully digital with transmit power <strong>of</strong> at least 15Kw.<br />

2 RF system should have maximum number <strong>of</strong> independent RF receiving channels with each having a band width <strong>of</strong><br />

1MHZ or more.<br />

3.Should have necessary hardware to support phased array coils.<br />

4 Specify frequency stability and amplifier resolution.<br />

5 RF system should be compatible with parallel imaging techniques. It should be able to support time reductions with<br />

compatible coils in 2D/3D imaging in Body/Neuro imaging upto atleast a factor <strong>of</strong> 2.<br />

6 RF amplifier should be solid state for over all better performance.<br />

7. Optical RF receiver system with RF receiver located on the magnet system inside the shielded scan room<br />

8. RF transmit technology-Multi drive RF transmit system should be <strong>of</strong>fered with two independent output channels to<br />

improve RF uniformity and signal homogeniety.<br />

4 RF Coils<br />

RF coils form an important part <strong>of</strong> imaging chain. High density posteriorly embedded element coils<br />

compatible with parallel imaging techniques should be quoted. All the array coils should be compatible with<br />

Parallel Imaging techniques.<br />

1 The main body coil integrated to the magnet must be Quadrature/CP. In addition to this following coil should be<br />

quoted.<br />

26