The latest model / version <strong>of</strong> CT Sacnner available at the time <strong>of</strong> shipment should be provided inspite <strong>of</strong><br />
final quoted configuration.<br />
Essential specifications <strong>of</strong> the proposed state <strong>of</strong> the art MDCT are given below.<br />
The specifications mentioned are the essential minimum requirements <strong>of</strong> the tender. While the terms and<br />
conditions are mentioned separately<br />
Description<br />
A State <strong>of</strong> Art Spiral Multi slices / Multi detector row CT Scanner capable <strong>of</strong> generating 128 slices per 360 0 rotation<br />
with isotropic micro voxel imaging to perform whole body scanning (including trauma, vascular, cardiac, pediatric,<br />
neurologic applications), the system should have at least 64 rows <strong>of</strong> detector with simultaneous generation <strong>of</strong> minimum<br />
<strong>of</strong> 128 or more slices,<br />
1 PATIENT TABLE<br />
a) Load Carrying Capacity 180 Kg or more with positional accuracy <strong>of</strong> +/- 0.25 mm or less<br />
b) Metal free scannable range <strong>of</strong> 150 cm or more<br />
c) Mention the horizontal movement range and it should be sufficient to cover the patient from head to toe.<br />
Also specify the horizontal speed<br />
d)Specify vertical range, vertical scannable range and elevation speed.<br />
e) Specify width and length <strong>of</strong> table – distance between gantry front and table base (min. & max.)<br />
f) Automatic re-centering facility to the scan iso centre in all three directions should be available.<br />
2 SCANNER GANTRY<br />
a) Aperture minimum 70 cms or higher<br />
b)Scan field <strong>of</strong> view in acquisition mode should be 50 cm. Specify minimum and maximum FOV with<br />
increment values<br />
c) Tilt : +/- 30 degree adjustable from gantry and console by remote<br />
d) 3-D laser lights for positioning. Isocenter in all planes should be maintained ever when the gantry tilts<br />
e) The gantry should be provided with user control panels on either side for easy positioning<br />
f) Specify focus to iso centre distance and focus to detector distance<br />
g) Specify the entire range <strong>of</strong> rotation time for 360 0 (lesser will be preferred).<br />
3 X-RAY GENERATOR<br />
a) High frequency type inverter based<br />
b) Power output : 70 KW or more at 120 Kvp<br />
c) mA range : 100– 600 mA or more<br />
d) Kv range 80 to 130 or more.<br />
e) Specify the max. mA allowed for 120 Kv for how long <br />
4 X-RAY TUBE<br />
a) Tube Voltage : 80 - 140 kVp or more<br />
b) Anode Heat Storage capacity <strong>of</strong> at least 7.5 MHU or more. Alternatively the tube should have a high<br />
dissipation rate <strong>of</strong> 1600 KHU/Min or more. Specify the dissipation rate also. And mention the details <strong>of</strong><br />
cooling system <strong>of</strong> the tube.<br />
c) Specify the focal spot size as per IEC standard.<br />
d) Source output <strong>of</strong> 70 KW or more.<br />
5 SPIRAL ACQUISITION<br />
a) Scan time should be 0.35 sec or less, for full 360 deg rotation<br />
b) Specify the time in minutes for warm up from fully <strong>of</strong>f in normal as well as in emergency situation.<br />
c) Minimum slice thickness should be 0.625 mm or less. Specify minimum slice thickness in axial, spiral and<br />
cardiac modes.<br />
32