The latest model / version <strong>of</strong> CT Sacnner available at the time <strong>of</strong> shipment should be provided inspite <strong>of</strong><br />

final quoted configuration.<br />

Essential specifications <strong>of</strong> the proposed state <strong>of</strong> the art MDCT are given below.<br />

The specifications mentioned are the essential minimum requirements <strong>of</strong> the tender. While the terms and<br />

conditions are mentioned separately<br />

Description<br />

A State <strong>of</strong> Art Spiral Multi slices / Multi detector row CT Scanner capable <strong>of</strong> generating 128 slices per 360 0 rotation<br />

with isotropic micro voxel imaging to perform whole body scanning (including trauma, vascular, cardiac, pediatric,<br />

neurologic applications), the system should have at least 64 rows <strong>of</strong> detector with simultaneous generation <strong>of</strong> minimum<br />

<strong>of</strong> 128 or more slices,<br />

1 PATIENT TABLE<br />

a) Load Carrying Capacity 180 Kg or more with positional accuracy <strong>of</strong> +/- 0.25 mm or less<br />

b) Metal free scannable range <strong>of</strong> 150 cm or more<br />

c) Mention the horizontal movement range and it should be sufficient to cover the patient from head to toe.<br />

Also specify the horizontal speed<br />

d)Specify vertical range, vertical scannable range and elevation speed.<br />

e) Specify width and length <strong>of</strong> table – distance between gantry front and table base (min. & max.)<br />

f) Automatic re-centering facility to the scan iso centre in all three directions should be available.<br />

2 SCANNER GANTRY<br />

a) Aperture minimum 70 cms or higher<br />

b)Scan field <strong>of</strong> view in acquisition mode should be 50 cm. Specify minimum and maximum FOV with<br />

increment values<br />

c) Tilt : +/- 30 degree adjustable from gantry and console by remote<br />

d) 3-D laser lights for positioning. Isocenter in all planes should be maintained ever when the gantry tilts<br />

e) The gantry should be provided with user control panels on either side for easy positioning<br />

f) Specify focus to iso centre distance and focus to detector distance<br />

g) Specify the entire range <strong>of</strong> rotation time for 360 0 (lesser will be preferred).<br />

3 X-RAY GENERATOR<br />

a) High frequency type inverter based<br />

b) Power output : 70 KW or more at 120 Kvp<br />

c) mA range : 100– 600 mA or more<br />

d) Kv range 80 to 130 or more.<br />

e) Specify the max. mA allowed for 120 Kv for how long <br />

4 X-RAY TUBE<br />

a) Tube Voltage : 80 - 140 kVp or more<br />

b) Anode Heat Storage capacity <strong>of</strong> at least 7.5 MHU or more. Alternatively the tube should have a high<br />

dissipation rate <strong>of</strong> 1600 KHU/Min or more. Specify the dissipation rate also. And mention the details <strong>of</strong><br />

cooling system <strong>of</strong> the tube.<br />

c) Specify the focal spot size as per IEC standard.<br />

d) Source output <strong>of</strong> 70 KW or more.<br />

5 SPIRAL ACQUISITION<br />

a) Scan time should be 0.35 sec or less, for full 360 deg rotation<br />

b) Specify the time in minutes for warm up from fully <strong>of</strong>f in normal as well as in emergency situation.<br />

c) Minimum slice thickness should be 0.625 mm or less. Specify minimum slice thickness in axial, spiral and<br />

cardiac modes.<br />