7 ERROR MESSAGES<br />
Because of wrong usage or some error conditions <strong>LCA</strong> device displays the following messages:<br />
ERROR-3<br />
"ZERO LIMIT"<br />
Cannot perform Zeroing. This value cannot be set to zero by pressing Zero ( ) key. Because, there<br />
is a limitation for zeroing with maximum of %4. Can be done with Zero Calibration.<br />
ERROR-50<br />
ERROR-1<br />
"CALIB.KEY!"<br />
Calibration key is disabled. To perform this operation, modification should be done on the <strong>LCA</strong> board.<br />
"MAXIMUM!!!"<br />
Screen value is over. (Over than Max+9e, Max:Maximum capacity, e: step value)<br />
ERROR-22<br />
ERROR-90<br />
ERROR-5<br />
ERROR-91<br />
"Eeprom Mem"<br />
Non-volatile memory error. Please call service.<br />
"CONV.TOUT!"<br />
ADC conversion is timed out. Chip not responding. Please call service.<br />
"ADC RANGE!"<br />
Internal counts are in over state. Check ADC gain setup.<br />
"TEMP.TOUT!"<br />
Time out for temperature sensor. Not responding. Call service.<br />
ERROR-99 "WATCHDOG! "<br />
There is an unknown soft lock occurred and device automatically RESET itself.<br />