Washington Closure Hantford<br />
June 17, 2010<br />
The requested analyses were:<br />
Alpha Spectroscopy<br />
Plutonium-238, -239/240 by method RL-ALP-002 (RICH-<strong>RC</strong>-5O 10)*<br />
Gas Proportional Counting<br />
Strontium-90 by method RL-GPC-003 (RICll-<strong>RC</strong>-5O06)*<br />
Gamma Spectroscopy<br />
Gamma Spec by method RL-GAM-001I (RJCII-<strong>RC</strong>-5017)*<br />
Liquid Scintillation Counting<br />
Tritium by method RL-LSC-005 (RICI I<strong>RC</strong>-5007)*<br />
Carbon-14 by method RL-LSC-008 (RICH-<strong>RC</strong>-5022)*<br />
Nickel-63 by method RL-LCS-017 (RICHl-<strong>RC</strong>-5069)*<br />
*SOPj ID's changed effective 7-01-2008. Attached is a cross reference until SOP ID's are changed in all systems.<br />
IV. Quality Control<br />
Thc analytical results for each analysis performed includes a minimum of one laboratory control sample<br />
(LCS), one method (reagent) blank, and one duplicate sample analysis. Any exceptions have been noted<br />
in the "Comments" section.<br />
QC and sample results are reported in the same units.<br />
V. Comments<br />
Alphia Spectroscopy<br />
Plutonium-238. -239/240 by method RL-ALP-002 (RICH-<strong>RC</strong>-50 10):<br />
'The LCS, batch blank, samples and sample duplicate (JI9YB6) results are within contractual<br />
requirements.<br />
Gas Proportional Counting<br />
Strontium-90 by method RL-QPC-003 (RICH-<strong>RC</strong>-5006):<br />
'[he L.CS, batch blank, samples and sample duplicate (J I9Y-B5) results are within contractual<br />
requirements.<br />
Gamma Spectroscopy<br />
Gamma Spec by method RL-GAM-00lI (RICI I-<strong>RC</strong>-50 17):<br />
There was insufficient volume for a duplicate. Sample JI19YB34 was recounted on a different detector for<br />
the duplicate (J19YB4 DUP).<br />
The count times were reduced to meet the priority turn around time. The following samples and the LCS<br />
do not meet the CRDL for Eu-152, Eu-154 & Eu-155 due to reduced count times.<br />
JI9YB4 Jl9Y136 JI9YB7<br />
Except as noted, the LCS, batch blank, samples and sample duplicate (JI9YB4) results are within<br />
contractual requirements.<br />
TestAmerica Page 3 of 64