Configuring the <strong>TraumaCad</strong> Server<br />
<strong>TraumaCad</strong> Administrator’s <strong>Guide</strong><br />
NOTE:<br />
This is a cross organizational configuration, meaning that all the<br />
workstations are configured the same way without the ability to change the<br />
default settings from the client side. Any selection made here affect all the<br />
workstations and causes them to operate according to these settings.<br />
• Client inactivity timeout: If <strong>TraumaCad</strong> client is open, but is not<br />
currently being worked on (is dormant), a timeout occurs after the<br />
designated period of time specified in this field and the software<br />
closes down. It is important to note that all work performed on<br />
the client up to the time of the timeout is lost without the ability<br />
to restore it. This is a security feature.<br />
• Fuji Integration Version [Optional]: This only applies if you are<br />
working with a Fuji Synapse PACS system. (If you are using a<br />
different PACS system, then leave the default value at version 3.2).<br />
Due to various changes made in the Fuji integration services, you<br />
must specify your Fuji Synapse version. Only version 3.1 and 3.2 are<br />
supported. (The minor version, which is the third set of digits in a<br />
version number, is not relevant.)<br />
• Regex [Optional]: This option is specific to non-English,<br />
Fuji-integrated systems. Consult the Orthocrat Global Support<br />
Center before applying this setting if your server works with a<br />
non-English, Fuji system.<br />
5 Click Save Settings to save all these settings to the <strong>TraumaCad</strong> server<br />
and a confirmation window is displayed.<br />
If you have not been able to query/retrieve or commit back to the PACS<br />
system and you have checked that the configuration on both sides is<br />
correct, then please click the Restart Dicom Proxy button. This will restart<br />
<strong>TraumaCad</strong> Server’s DICOM component.