Localization <strong>and</strong> <strong>Biopsy</strong><br />
Process<br />
Needle Penetration at Incorrect Location<br />
The mamma must be correctly immobilized.<br />
If the mamma is not properly immobilized, then it may slip out<br />
of position, <strong>and</strong> the data recorded by the <strong>MR</strong>I are then incorrect.<br />
• Ensure that as much mamma tissue as possible is held between the slat lateral<br />
plate <strong>and</strong> the compression plate.<br />
• Insert the marker block supplied into the compression plate.<br />
1<br />
Posterior<br />
Medial Lateral<br />
Anterior<br />
• Detect the marker block (1) <strong>and</strong> the lesion (2) in the <strong>MR</strong>I using sagittal Slices.<br />
• Mark the lesion with a permanent auxiliary line from the software toolbox of your<br />
<strong>MR</strong>I device. Go back to the slice where you can see the marker block <strong>and</strong> the<br />
impression of the grid.<br />