Localization <strong>and</strong> <strong>Biopsy</strong><br />

Process<br />

Needle Penetration at Incorrect Location<br />

The mamma must be correctly immobilized.<br />

If the mamma is not properly immobilized, then it may slip out<br />

of position, <strong>and</strong> the data recorded by the <strong>MR</strong>I are then incorrect.<br />

• Ensure that as much mamma tissue as possible is held between the slat lateral<br />

plate <strong>and</strong> the compression plate.<br />

• Insert the marker block supplied into the compression plate.<br />

1<br />

Posterior<br />

Medial Lateral<br />

Anterior<br />

• Detect the marker block (1) <strong>and</strong> the lesion (2) in the <strong>MR</strong>I using sagittal Slices.<br />

• Mark the lesion with a permanent auxiliary line from the software toolbox of your<br />

<strong>MR</strong>I device. Go back to the slice where you can see the marker block <strong>and</strong> the<br />

impression of the grid.<br />

<strong>MR</strong>-<strong>BI320</strong>-<strong>PA</strong><br />

6<br />

47