The <strong>Immobilization</strong> Device can be rotated by 60°. This ensures optimum access<br />

to the lesion (ca. 270°). For guidance, a telescopic assembly, the post <strong>and</strong> pillar<br />

system, provides access also to regions very close to the chest wall (axilla). An<br />

additional access possibility is offered by the alternative use of compression plates<br />

<strong>and</strong> needle blocks. <strong>Biopsy</strong> access is single-sided medial, single-sided/double-sided<br />

lateral or single-sided/double-sided cranio-caudal.<br />

Safety Concept<br />

Operating Principle<br />

The following safety concept must be strictly followed while performing a biopsy.<br />

A control scan must be made after each work step to verify the correct completion<br />

of the previous work steps to be able to recognize incorrect adjustments of the<br />

needle guides bases <strong>and</strong> needle blocks during the biopsy. That is the only way<br />

incorrect adjustments, incorrectly read scale values <strong>and</strong> other operator errors can<br />

be recognized in order to prevent injuries to the patient.<br />

The control scan procedure is described in Chapter “Device Use”, in the description<br />

of the biopsy procedure, <strong>and</strong> the correct result illustrated in the text.<br />

<strong>MR</strong>-<strong>BI320</strong>-<strong>PA</strong>