STANDARD OPERATING PROCEDURES FOR SURFACE WATER QUALITY SAMPLINGFIGURE 9.2. Pebble count excel sheet.9.3 POST-TRIP CALIBRATIONSCalibrate pH, dissolved oxygen and conductivity when you return to the office. Values shouldagree with the precalibration results within the following limits.Parameter Acceptable RangepHPost-trip reading should be within 0.3 standard units of the pre-trip readings <strong>for</strong>each buffer solution used. Hydrolab instrument specifications specify aninstrument tolerance of +/- 0.2 SU, and additional 0.1 SU is permitted <strong>for</strong> minorenvironmental differences (temperature, etc.).Conductivity EC readings shall be within 10% of pre-trip post-calibration values.Dissolved Dissolved oxygen saturation values will be evaluated site by site according toOxygen calibration per<strong>for</strong>mance in the field. Readings will be evaluated after sitecalibration by stabilization time (should be within 2 minutes) and adherence to afull 100% calibration <strong>for</strong> YSI. Hydrolab measurements should be within 10%.TABLE 9.1 Post-trip calibrations.REVISED SEPTEMBER 2012 180