10250Take-Apart Laparoscopic Instruments987240230220210200190180Directions for Disassembly and Re-Assembly<strong>Integra</strong> Jarit Take-Apart Laparoscopic Instruments are reusable, non-sterile, fully-assembled, monopolar electrosurgicalinstruments comprised of three individual parts – handle, shaft and stainless steel insert. Disassembly is completed ina few short steps; however, inserts must always be disengaged prior to removing the shaft from the handle or damagemay result. For re-assembly, the shaft must be connected to the handle for the insert to be seated fully within thehandle assembly.A. DISASSEMBLYStep 1:For handles with ratchets: Push gold-plated lever downto detach/disassemble ratchet connection. For handleswithout ratchets: Go directly to Step 2.17065160150140130120Step 2:Note: Insert should be disengaged from the handleprior to unscrewing the shaft or damage to the handlemay result.Press and hold rounded button on the top leg of thehandle. Open handle by lifting until an audible pop isheard and push button is visibly tucked underneath thehandle assembly.110410090Step 3:Insert is now disengaged from handle and may beremoved by grasping distal end and pulling from shaft.38070602150403020Step 4:For handles with rotation: Hold rotating wheel inplace, loosen screw connection nut, then separateshaft from handle. For handles without rotation:Hold handle in place, loosen screw connection nut,then separate shaft from handle.INCHES10MILLIMETERS34 877-468-5572 USA 800-275-7192 fax integralife.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!