Unit Operation Instruction<strong>KilnLoq</strong> <strong>probe</strong> & EXDBlow Back hoses (Cleaning Air)Cooling water connection:Flange for mounting <strong>probe</strong>10 mm stainless steel SwagelokISO 7/1-Rp-1 Union(1”female pipe tread)See Drawing “<strong>KilnLoq</strong> <strong>probe</strong>” <strong>and</strong> EXD2Mating Flange is NOT supplied by FLSmidth A/S3.2 TECHNICAL INFORMATION KILNLOQ 400 PROBEDimensions:LengthHeight (filter house cover):Width (filter house cover):Sample end length:Available length for insertion in process:Diameter in process:Sample Filter:MaterialFilter sizeNOTE: Lengths depending on specific sampleend length.2000-4000mm400mm230mm1500/2500/3000/3700 mm1000/2000/2500/3200 mm (from EXD processflange)90 mmAISI 316L20µm ,102xØ50/40Process conditions:Measuring points:Kiln inlet or similarTemperature Max. 400°CDust load Max. 1000 g/m 3Weight:Weight, dry:150 kgTemperatures:Ambient temperature: -10°C to 55°C 2Storage temperature: -15°C to 60°CEquipment must be protected against heatradiationPower supply:Power supply from GMU:380-460 VAC, 2,5 kWPlease see “System Data sheet” for specificproject.2 Probe must be protected to prevent condensation <strong>and</strong> freezing.Page 8 of 27