Portable Trace Oxygen AnalyzerOperationalarm. A HI alarm activates when the oxygen concentration rises abovethe setpoint. The LO alarm activates when the oxygen concentrationfalls below the setpoint.To set the alarm as either a HI or LO alarm:1. Use the UP/DOWN keys to navigate to the HI-LO ALARMscreen. Then press ENTER to enter the setup screen.2. Use the UP/DOWN keys to toggle between HI and LOoption. Then press ENTER to accept the displayed status.3. Press ENTER again to save the alarm status and return to theHI-LO ALARM screen.3.2.9 LTCH ALARM ScreenThe alarm on the <strong>Model</strong> <strong>3110</strong> can be set up as latching or nonlatching.A latching alarm, once triggered, will remain in alarm statusuntil recognized and reset. Even if the concentration changes back to anon-alarm concentration, a latched alarm will remain in alarm status. Anon-latching alarm will cease to alarm when the concentration falls orrises to a non-alarm value.To set the alarm to either latching or non-latching:1. Use the UP/DOWN keys to navigate to the LTCH ALARMscreen. Then press ENTER to enter the setup screen.2. Use the UP/DOWN keys to toggle between LTCH and NON-LTCH options. Then press ENTER to accept the displayedstatus.3. Press ENTER again to save the alarm status and return to theLTCH ALARM screen.<strong>Teledyne</strong> <strong>Analytical</strong> <strong>Instruments</strong> 31