Condition Cause Corrective Action1 - ERR • The calibration weight is less than 1/2full capacity2-ERR • The data error exceeds 1%• Defective weighing mechanism• Reinitiate calibration with a test mass equalto the nominal capacity of the balance• Consult your local <strong>Rice</strong> <strong>Lake</strong> <strong>Weighing</strong><strong>Systems</strong> distributor• Consult your local <strong>Rice</strong> <strong>Lake</strong> <strong>Weighing</strong><strong>Systems</strong> distributor3-ERR4-ERR• Calibration has started with a load onthe platter• Remove load and reinitiate calibrationprocess• Consult your local <strong>Rice</strong> <strong>Lake</strong> <strong>Weighing</strong><strong>Systems</strong> distributorU - ERR • Material wedged underneath theplatform support plate• Remove platform and platform supportplate, then clean and reassembleTable 6-1. Troubleshooting Chart (Continued)26 <strong>QB</strong> <strong>Series</strong> Operation Manual