12.07.2015 Views

Recalibration of Paroscientific Transducers ... - Paroscientific, Inc.

Recalibration of Paroscientific Transducers ... - Paroscientific, Inc.

Recalibration of Paroscientific Transducers ... - Paroscientific, Inc.

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Paroscientific</strong>, <strong>Inc</strong>. Procedure Document: 8140-001Title: <strong>Recalibration</strong> <strong>of</strong> <strong>Paroscientific</strong> <strong>Transducers</strong>Revision: ANote that oil-filled units may not be pressurized with pneumatic (gas filled) pressurelines. A complete conformance check requires a hydraulic primary standard.Note that air-filled pressure units need additional time for pressurestabilization, since the expansion and contraction <strong>of</strong> gas induces adiabatic cooling orheating. After full scale pressure changes above a few hundred psia, the transducershould be stabilized for at least 5 minutes before taking dat11 <strong>of</strong> 13

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!