<strong>Advanced</strong> Instruments Inc.3.3 <strong>PPM</strong>AUTO SAMPLING10 <strong>PPM</strong> RANGE24.5 C 100 KPALO1 2<strong>PPM</strong>10<strong>PPM</strong> HI2Bypass Alarms:This feature, separate from CONFIG ALARMS above, enables the user to disable the local audible alarm and relay contactsduring calibration or servicing. The alarms are enabled when the alarm condition is corrected.1. Access the MAIN MENU by pressing the MENU key.2. Advance the reverse shade cursor using the ARROW keys to highlight BYPASS ALARMS.3. The following displays appears:MAIN MENUAUTO SAMPLEMANUAL SAMPLECALIBRATIONCONFIG ALARMSBYPASS ALARMS4. Press the ENTER key to bypass and disable both the local audible alarm and relay contacts. After 3 seconds, the systemreturns to SAMPLING mode.5. Note: The appropriate alarm setting will alternately reverse shades indicating the alarm condition exists but the BYPASSALARMS feature has disabled the local audible alarm and relay contact. The alarms are enabled when the alarm condition iscorrected.3.3 <strong>PPM</strong>AUTO SAMPLING10 <strong>PPM</strong> RANGE24.5 C 100 KPALO1 2<strong>PPM</strong>10<strong>PPM</strong> HI2Start-Up is complete . . . proceed to Calibration21