Troubleshooting GuideThe <strong>2200</strong> <strong>Enteral</strong> <strong>Feeding</strong> <strong>Pump</strong> is a reliable, electromechanical device.As with any electromechanical device, problems may occur. The followingtips will help you correct minor problems.SITUATIONLOW BATalarm activated.OCCL/EMPTYalarm activated.POSSIBLE CAUSEBattery charge is belowperformance level.<strong>Feeding</strong> container is emptyOcclusion is restrictingflow.Drip chamber walls coatedwith formula.Drip chamber not properlyplaced in pump.Too much fluid in dripchamber.Drop sensors need cleaning.<strong>Pump</strong> is on hold for morethan 2-1/2 minutes.SOLUTIONPlug cord into AC outlet.Replace or refill feedingcontainer.Find point of occlusion:kinked tubing or closedroller clamp in pump set;feeding tube; container, andcorrect.Replace pump set.Review operating instructionsand position properly.Replace pump set.Clean with cotton swabdampened with isopropylalcohol.Press RUN/HOLD button toreturn to hold mode.16