Troubleshooting Guide (Cont’d)SITUATIONRate changealarm activated.Dose completealarm activated.Free flowalarm activated.POSSIBLE CAUSEINC, DEC, and TOTALRESET may only bepressed when pump ison hold.<strong>Pump</strong> has deliveredpreset volume.<strong>Pump</strong> set hasbecome dislodged.SOLUTIONPress RUN/HOLD buttononce to stop pump.Change rate. PressRUN/HOLD to restartpump.Press total reset. PressRUN/HOLD. If additionalvolume is needed,pump will operate untildose is completedagain.Press RUN/HOLD buttononce to stop pump.Correct set placementon rotor.Drip chamber walls coatedwith formula.Replace pump set.17