InstallationThreshold Alarm 2:Can be configured as high (actuates when concentration is abovethreshold), or low (actuates when concentration is below threshold).Configured as failsafe or non-failsafeConfigured as non-latchingCan be configured out (defeated).Can be set to trigger by levels of CH4, NMH, or THC.System Alarm:Actuates when power is removed from <strong>analyzer</strong>.It is triggered if <strong>4040</strong> fails to ignite after five times.Note: (Reset by pressing the STANDBY button to remove power.Then press STANDBY again and any other button exceptSYSTEM to resume. Further detail can be found in Chapter 4,Section 4.5.15.Table 3-4: Alarm Relay Contact PinsPin Contact45 Threshold Alarm 1, normally closed contact28 Threshold Alarm 1, moving contact46 Threshold Alarm 1, normally open contact42 Threshold Alarm 2, normally closed contact44 Threshold Alarm 2, moving contact43 Threshold Alarm 2, normally open contact36 System Alarm, normally closed contact20 System Alarm, moving contact37 System Alarm, normally open contact<strong>Teledyne</strong> <strong>Analytical</strong> Instruments 31