12.07.2015 Views

PhD6 Multi-Gas Detector Reference Manual - Honeywell Analytics

PhD6 Multi-Gas Detector Reference Manual - Honeywell Analytics

PhD6 Multi-Gas Detector Reference Manual - Honeywell Analytics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Fresh air/zero calibrationsmay only be performed in an atmosphere thatis known to contain 20.9% oxygen, 0.0% LELand 0 PPM toxic gas.To initiate the fresh air/zero calibration:1. Press the MODE button three times withintwo seconds to begin the fresh air/zerocalibration sequence. The PHD6 will brieflydisplay AUTO CAL and then begin a 5-second countdown.2. Press the MODE button before the end of the5-second countdown to begin the freshair/zero calibration. The fresh air/zerocalibration is initiated when the PHD6 shows“Calibrating” on the screen.→3. The PHD6 will indicate when the freshair/zero calibration is complete. It will thenproceed to a second 5-second countdown forthe gas calibration. If gas calibration is notrequired, allow the countdown to reach 0without pressing the MODE button.→For instructions on the <strong>Gas</strong> Calibration,proceed to section 4.3.4.2.1 Fresh air calibration failureIn the event of a fresh aircalibration failure, thealarms will be activated andthe instrument will displaythe following screen. Notethat the sensor(s) that failthe zero calibration areshown (in this case, CO)After 3 seconds, the PHD6will return to the current gas readings screen andthe visual and audible alarms will cease.When calibration is due, the triangular warningsymbol along with the span bottle icon thePHD6’s status bar will showIf a successful fresh air calibration is notperformed prior to instrument shut down, thePHD6 will note that Fresh Air Calibration is dueduring instrument start up.Possible causes and solutions1. The atmosphere in which the instrument islocated is contaminated (or wascontaminated at the time the instrument waslast fresh air calibrated.2. A new sensor has just been installed.3. Instrument has been dropped or bangedsince last turned on.4. There has been a significant change intemperature since the instrument was lastused.Recommended action:Take the instrument to fresh air and allowreadings to stabilize. Perform the fresh air/zeroadjustment again. If the manual fresh air/zeroprocedure fails to correct the problem, performthe manual fresh air / zero calibration procedureas described in section 4.2.2 below.4.2.2 Forced fresh air calibrationThe PHD6 includes safeguards to prevent freshair calibration in contaminated environments. Ifthe standard fresh air calibration fails a secondtime, the instrument may be “forced” to acceptthe fresh air calibration by performing the manualfresh air calibration.Fresh air calibrations mayonly be performed in an atmosphere that isknown to contain 20.9% oxygen, 0.0% LELand 0 PPM toxic gas. Performing a fresh aircalibration in a contaminated atmospheremay lead to inaccurate and potentiallydangerous readings.1. Initiate the standard fresh air / zerocalibration sequence by pressing the MODEbutton three times in rapid succession. The5-second countdown will begin.2. Press and hold the down arrow key and thenpress the MODE button before the end of the5-second countdown. Continue to hold thedown arrow.3. The fresh air/zero calibration is completewhen the instrument begins another 5-second countdown for the gas calibration. Ifgas calibration is not required, allow thecountdown to reach 0 without pressing theMODE button.If the PHD6 still fails to calibrate after thisprocedure is attempted, contact <strong>Honeywell</strong><strong>Analytics</strong>.4.2.3 Fresh air calibration in a contaminatedatmosphereTo fresh air calibrate the PHD6 in a contaminatedatmosphere, connect a cylinder of “zero air”containing 20.9% oxygen and no contaminants tothe PHD6 and flow gas to the instrument. Then20

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!