6.6.1 Replacing pump filters1. Remove the two screws that hold the inletport to the pump.2. Gently pull the dust filter holder free of thepump.3. Remove and replace the dust filter that islocated in the holder.4. The hydrophobic filter is located beneath theinlet port in the pump housing. Use a smallscrewdriver or other object to punch throughthe filter and remove it. The gasket that sitsbetween the inlet port and the filter shouldcome out with the filter.5. Place the new hydrophobic filter with the filterside down in place of the one removed instep 4. The gasket should be located on topof the filter and should sit against the dustfilter holder, which will be reinstalled in step6.6. Replace the dust filter holder (which now hasa new filter in it) and secure it with the twoscrews removed in step 1.30