92 A FIRST APPENDIX<br />

trons than a standard PMT with bialkali pho<strong>to</strong>cathode [42]. The collaboration<br />

reported in [55] the resolution of a 8x8x30 cm 3 (8x8x23 cm 3 )<br />

crystal coupled <strong>to</strong> two PMTs (9269A from Electron Tube Ldt., PbCs pho<strong>to</strong>cathode,<br />

quartz window) as 3.5 % (4.6 %) RMS at 662 keV with a pho<strong>to</strong>electron<br />

yield of 5.6 (4.5) pe/keV. Kim et al. [43] report further resolution<br />

(RMS at 662 keV) ranging from 4.0 <strong>to</strong> 5.6 % of crystal with a length<br />

of 30 cm and front and rear surfaces ranging from 55x55 <strong>to</strong> 70x70 cm 2 ,<br />

using two 3-in Pmts (ET D726UK).<br />

Renker [60] reported energy resolutions of scintillation detec<strong>to</strong>rs with<br />

APD <strong>for</strong> LSO crystals of 12.1 % FWHM, 22.4 % FWHM <strong>for</strong> BGO at<br />

511 keV and 4.3 % FWHM with YAP:Ce crystals at 662 keV.<br />

Ikagawa et al. [61] CsI:Tl 10x10x10 mm 3 with Hamamatsu APD S8664-<br />

1010N (10x10 mm 2 ) reached 4.9 % FWHM at 662 keV in comparison<br />

with 5.9 % (FWHM at 662 keV) with the PMT R7899EG (Hamamatsu,<br />

1 inch diameter). 28 primary pe/keV were produced in APD. Ikagawa<br />

et al. [61] reported that APD can efficiently detect weak scintillation<br />

light of big crystals, even if the scintilla<strong>to</strong>r and APD size do not match.<br />

They couple an APD via a light guide <strong>to</strong> the rear face of a 48x3x300 mm 3<br />

big BGO plate and reached a energy resolution of 20.9 % (FWHM at<br />

662 keV and -20 ◦ ) in comparison <strong>to</strong> 7.1 % (FWHM at 662 keV and -20 ◦ )<br />

<strong>for</strong> a 10x10x10 mm APD and an equal APD. At room temperature (20 ◦ )<br />

the APD readout of the small BGO crystals showed a resolution of 8.3 %<br />

and 10.4 % <strong>for</strong> the PMT readout (FWHM at 662 keV). Asano et al. [62]<br />

CWO crystal (5x5x5 mm 3 ) with APD (Hamamatsu S8664-55, 5x5 mm 2 )<br />

and Hamamatsu PMT (R6231HA). Best results were obtained at -30 ◦ C<br />

with 6.4 and 7.0 % (FWHM at 662 keV). At 20 ◦ the APD and PMT had<br />

resolutions of aro<strong>und</strong> 9.2 and 7.6 % (FWHM at 662 keV).