13.07.2015 Views

Instructions for cleaning and sterilization of Humeca medical ...

Instructions for cleaning and sterilization of Humeca medical ...

Instructions for cleaning and sterilization of Humeca medical ...

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

3.6 InspectionAll <strong>medical</strong> devices must be inspected be<strong>for</strong>e preparation <strong>for</strong> sterilisation.Visual inspection without magnifying glass under good lighting conditions will generally suffice. All parts <strong>of</strong> the devices must be checked<strong>for</strong> visible soil, damage <strong>and</strong>/or corrosion. Special attention should be paid to:• Places where soil “continues to stick”, such as interconnecting surfaces, hinges, shafts, etc.• Recesses (holes, etc.);• Elements where soil can be pressed against the device;• Check <strong>for</strong> damage to the extent that the device no longer works or blemishes are <strong>for</strong>med that can damage tissue or surgicalgloves.Function checks should be per<strong>for</strong>med where possible:• A check should be made to see whether interconnecting devices have been properly assembled;• Medical devices with moving parts should be activated to check whether they still work properly (a lubricant <strong>for</strong> <strong>medical</strong>applications can be used that is suitable <strong>for</strong> steam sterilisation if required <strong>and</strong> only according to the instructions <strong>for</strong> use).4. SterilisationSteam sterilisation is recommended. This is suitable to sterilise <strong>Humeca</strong> <strong>medical</strong> instruments. The design <strong>and</strong> per<strong>for</strong>mance <strong>of</strong> theautoclave, however, may affect the efficiency <strong>of</strong> the process. Healthcare institutions should validate their procedure using the materials <strong>and</strong>the operators who usually h<strong>and</strong>le the instruments.4.1 Preparation be<strong>for</strong>e sterilisationUse special containers from <strong>Humeca</strong> where applicable.4.2 Sterilisation processSteam sterilisation (moist heat) in an autoclave using a pre-vacuum (<strong>for</strong>ced air) is recommended. Autoclaves must satisfy therequirements <strong>of</strong> <strong>and</strong> must be validated in accordance with ISO 13683.<strong>Humeca</strong> instruments has validated autoclave cycles <strong>for</strong> the sterilisation <strong>of</strong> complete <strong>medical</strong> devices.The following process parameters are validated by <strong>Humeca</strong> instruments <strong>and</strong> recommended <strong>for</strong> sterilisation.Method: steam sterilisation in accordance with ANSI/AAMI/ISO 11134.Cycle:Exposure time 1 :Temperature:Pressure:Drying time 2 :saturated steam with vacuum / fractional <strong>for</strong>ced air.4 minutes.The exposure time can be extended to 18 minutes in order to comply with the recommendations from WHO,RKI, etc. <strong>Humeca</strong> <strong>medical</strong> instruments can withst<strong>and</strong> such sterilisation cycles.134 °C - 137 °C (273 °F - 279 °F).2 bar (30 PSI) or 3 bar (45 PSI).Varies depending on the load, packaging method <strong>and</strong> materials. A drying time <strong>of</strong> 8 minutes is required <strong>for</strong> theelectric devices <strong>for</strong> heavy use <strong>of</strong> <strong>Humeca</strong> instruments.1Exposure time: period in which the load <strong>and</strong> the complete chamber are kept at the sterilisation temperature.2Drying time: the period in which steam is removed from the chamber <strong>and</strong> the chamber pressure is reduced to let the condensate <strong>of</strong> theload evaporate, either through sustained drainage or through the injection <strong>and</strong> extraction <strong>of</strong> hot air or other gases.REMARK: final responsibility <strong>for</strong> the validation <strong>of</strong> sterilisation techniques <strong>and</strong> equipment lies with the healthcare institution.To guarantee optimal treatment, all cycles <strong>and</strong> methods must be validated <strong>for</strong> different sterilisation chambers, packaging methods <strong>and</strong>/orall sorts <strong>of</strong> configurations <strong>for</strong> the load.6

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!