Instructions for cleaning and sterilization of Humeca medical ...
Instructions for cleaning and sterilization of Humeca medical ...
Instructions for cleaning and sterilization of Humeca medical ...
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
3.6 InspectionAll <strong>medical</strong> devices must be inspected be<strong>for</strong>e preparation <strong>for</strong> sterilisation.Visual inspection without magnifying glass under good lighting conditions will generally suffice. All parts <strong>of</strong> the devices must be checked<strong>for</strong> visible soil, damage <strong>and</strong>/or corrosion. Special attention should be paid to:• Places where soil “continues to stick”, such as interconnecting surfaces, hinges, shafts, etc.• Recesses (holes, etc.);• Elements where soil can be pressed against the device;• Check <strong>for</strong> damage to the extent that the device no longer works or blemishes are <strong>for</strong>med that can damage tissue or surgicalgloves.Function checks should be per<strong>for</strong>med where possible:• A check should be made to see whether interconnecting devices have been properly assembled;• Medical devices with moving parts should be activated to check whether they still work properly (a lubricant <strong>for</strong> <strong>medical</strong>applications can be used that is suitable <strong>for</strong> steam sterilisation if required <strong>and</strong> only according to the instructions <strong>for</strong> use).4. SterilisationSteam sterilisation is recommended. This is suitable to sterilise <strong>Humeca</strong> <strong>medical</strong> instruments. The design <strong>and</strong> per<strong>for</strong>mance <strong>of</strong> theautoclave, however, may affect the efficiency <strong>of</strong> the process. Healthcare institutions should validate their procedure using the materials <strong>and</strong>the operators who usually h<strong>and</strong>le the instruments.4.1 Preparation be<strong>for</strong>e sterilisationUse special containers from <strong>Humeca</strong> where applicable.4.2 Sterilisation processSteam sterilisation (moist heat) in an autoclave using a pre-vacuum (<strong>for</strong>ced air) is recommended. Autoclaves must satisfy therequirements <strong>of</strong> <strong>and</strong> must be validated in accordance with ISO 13683.<strong>Humeca</strong> instruments has validated autoclave cycles <strong>for</strong> the sterilisation <strong>of</strong> complete <strong>medical</strong> devices.The following process parameters are validated by <strong>Humeca</strong> instruments <strong>and</strong> recommended <strong>for</strong> sterilisation.Method: steam sterilisation in accordance with ANSI/AAMI/ISO 11134.Cycle:Exposure time 1 :Temperature:Pressure:Drying time 2 :saturated steam with vacuum / fractional <strong>for</strong>ced air.4 minutes.The exposure time can be extended to 18 minutes in order to comply with the recommendations from WHO,RKI, etc. <strong>Humeca</strong> <strong>medical</strong> instruments can withst<strong>and</strong> such sterilisation cycles.134 °C - 137 °C (273 °F - 279 °F).2 bar (30 PSI) or 3 bar (45 PSI).Varies depending on the load, packaging method <strong>and</strong> materials. A drying time <strong>of</strong> 8 minutes is required <strong>for</strong> theelectric devices <strong>for</strong> heavy use <strong>of</strong> <strong>Humeca</strong> instruments.1Exposure time: period in which the load <strong>and</strong> the complete chamber are kept at the sterilisation temperature.2Drying time: the period in which steam is removed from the chamber <strong>and</strong> the chamber pressure is reduced to let the condensate <strong>of</strong> theload evaporate, either through sustained drainage or through the injection <strong>and</strong> extraction <strong>of</strong> hot air or other gases.REMARK: final responsibility <strong>for</strong> the validation <strong>of</strong> sterilisation techniques <strong>and</strong> equipment lies with the healthcare institution.To guarantee optimal treatment, all cycles <strong>and</strong> methods must be validated <strong>for</strong> different sterilisation chambers, packaging methods <strong>and</strong>/orall sorts <strong>of</strong> configurations <strong>for</strong> the load.6