13.07.2015 Views

Operating Instructions PPC-06/08-Plus v2009 (PDF - 1,3 MB) - Stauff

Operating Instructions PPC-06/08-Plus v2009 (PDF - 1,3 MB) - Stauff

Operating Instructions PPC-06/08-Plus v2009 (PDF - 1,3 MB) - Stauff

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

9.5. Memory function AUTO TRIGGERThe function Auto Trigger documents the process of recording measured values triggered bya defined start signal (for example pressure on channel 2 >> 125 bar). In response, asequence of measured values is automatically recorded until the previously setmeasurement time expires.Time-dependent functions (for example making operations or production cycles) aremeasured when recording measured values.The following parameters must be set:MenuTIME UNIT>Setting/Pre-selectionsechrsValueshNotesPre-selected time unit(trigger / pre-trigger)DURATION Number 100 Recording timePRETRIGGER> Number 1 Pre-trigger time(time before the trigger signal)TRIGGER> IN p1 Starting-point measuring channelLEVEL> Number 125 bar Start-point valueTRIG SLOPE> ▲▼ ▲ Ascending or descending edgeAUTO READY> YES, NO YES Recording of measured values isrepeated automatically37