Peak energy (keV)M.Sc <strong>Dissertation</strong>4.3 Energy calibration <strong>of</strong> MCABefore starting the experiment it was essential to calibrate the systems to findwhether it is linear or not. The system was calibrated using the spectrum collected fromthe 22 Na, 137 Cs and 60 Co standard sources with LaCl 3 :Ce(10%) detector. Each spectrumwas collected for 600 sec.Table 4.3 Energy calibration for the detectorSourcePeak Energy(keV)Channel No22 Na 511.00 135.46137 Cs 661.67 176.8160 Co 1173.24 313.6360 Co 1274.50 337.1822 Na 1332.50 356.17Using the values shown in Table 4.3 the MCA calibration graph was plotted, as shownin Fig 4.3.13501250y = 3.7479x + 1.528211501050950850750650550450120 160 200 240Peak Channel No280 320 360Figure 4.3 The MCA calibration graph <strong>of</strong> LaCl 3 :Ce(10%) gamma-ray detectorK. <strong>Suren</strong> 21