9 Control of Safety functions<br />
Test instructions<br />
Page 21 of 28<br />
<strong>User</strong> <strong>Manual</strong> BW3-070/2<br />
Self test at start<br />
Connect the Control Unit to mains supply and connect the Heating Pad. Check that the mains indicator LED<br />
blinks.<br />
Press Start/Standby and check that the following happens in the order described below.<br />
1. The selected mattress mode is displayed.<br />
2. Clicking from the safety relay can be heard<br />
3. The Control Unit symbol turns green<br />
4. The Heating Pad symbol turns green<br />
5. A short "beep" is emitted<br />
6. The system starts operating at 37 o C<br />
1) Self test in progress. The selected mattress mode is displayed (GEL<br />
or H2O). NOTE: Change of mode is now possible by pressing LEFT or<br />
RIGHT arrow, as indicated on the display. If the mattress mode is<br />
changed the unit will re-start the self test.<br />
2) Self Test Ready. The Control Unit and the Heating Pad symbol turns<br />
green<br />
A short "beep" is emitted and the red alarm LED flashes one time.<br />
The Control Units software version is displayed.<br />
Testing the uneven Heating Pad temperature alarm in H 2O mode.<br />
Start the system in H2O-mode, at 37°C without any mattress placed on the Heating Pad. The Alarm 12<br />
should be received within 15 minutes. (Note: Alarm 08 can sometimes occur as a consequence of the first<br />
alarm).<br />
Testing the uneven Heating Pad temperature alarm (GEL mode).<br />
Start the system in GEL-mode, at 37°C without any mattress placed on the heating Pad. The Alarm 14<br />
should be received within 15 minutes. (Note: Alarm 08 can sometimes occur as a consequence of the first<br />
alarm).<br />
Mains Power failure alarm<br />
This test is performed on a system with Mattress and a Heating Pad.<br />
1. Start the warming<br />
2. After about one minute, select a temperature different from 37 o C and make a note of it.<br />
3. Wait about 2 minutes and then disconnect the supply voltage by pulling out the mains plug.<br />
4. Check that the Control Unit alarm indicator starts to blink and that the acoustic alarm beeps<br />
intermittently for at least 10 minutes<br />
5. Switch on supply voltage again and check that the Power alarm is displayed. Press the alarm button.<br />
6. Restart and check that the warming starts at the set temperature selected before the power<br />
interruption.