144<br />
<strong>Vacuum</strong> gauges and controllers<br />
<strong>Vacuum</strong> controller<br />
<strong>Vacuum</strong> controller VNC 2<br />
PERFORMANCE FEATURES<br />
■ on-demand control of process vacuum, cooling<br />
water and venting; special program <strong>for</strong> operating<br />
vacuum networks like VACUU·LAN ®<br />
■ easy changing of vacuum set points also during<br />
operation<br />
automatic or manual setting of the pressure<br />
■ switching interval (hysteresis)<br />
integrated relay <strong>for</strong> switching pumps on/off<br />
■ (except VNC 2 VARIO)<br />
■<br />
The vacuum controller VNC 2 makes automatic, on-demand vacuum control possible either via two-point valve control<br />
or by switching the pump on or off. The version VNC 2 VARIO offers continuous vacuum control with VARIO ® pumps.<br />
The integrated alumina ceramic diaphragm vacuum sensor is chemically resistant, and offers outstanding measuring<br />
accuracy independent of gas type. The VNC 2 vacuum controller is designed <strong>for</strong> precise and reliable control<br />
of recurrent routine jobs that do not require regular adjustment. It works with components compatible with the<br />
VACUU·BUSTM system. Special versions of the controller, e.g., <strong>for</strong> mounting into laboratory furniture, are available.<br />
Further in<strong>for</strong>mation on request.<br />
selectable acoustic alert signals<br />
VNC 2<br />
TECHNICAL DATA<br />
Control range<br />
Measurement principle<br />
Measurement uncertainty<br />
Temperature coefficient<br />
<strong>Vacuum</strong> connection<br />
Permissible ambient temp. range storage / operation<br />
Max. media temp. <strong>for</strong> continuous operation / short times<br />
Material of outer housing<br />
Degree of protection<br />
Degree of protection of front side of built-in versions<br />
Rated mains voltage / mains frequency<br />
Max. power <strong>for</strong> VACUU·BUS TM (24V DC)<br />
Max. AC switching capacity ohmic(inductive) load below 30°C<br />
mbar / hPa<br />
Max. AC switching capacity ohmic(inductive) load at 30-40°C<br />
Dimensions (L x W x H)<br />
Weight<br />
Further in<strong>for</strong>mation at www.vacuubrand.com<br />
°C<br />
°C<br />
V / Hz<br />
mm<br />
kg<br />
ORDERING INFORMATION VNC 2<br />
100-230 V ~ 50-60 Hz IEC plug EN 60320 683070<br />
VNC 2 VARIO on request<br />
ACCESSORIES<br />
Rubber vacuum tubing DN 10 mm (686002)<br />
PTFE tubing DN 10/8 mm (638644)<br />
DKD calibration with first delivery (900216)<br />
DKD recalibration (900217)<br />
VACUU·BUS<br />
ITEMS SUPPLIED<br />
VNC 2<br />
1100 / 825 - 1 / 1<br />
Ceramic diaphragm (alumina), capacitive, gas indep., absolute pressure<br />
< +- 1 mbar/hPa/torr / +- 1 digit (after adjustment, constant temp.)<br />
< 0.07 mbar/hPa/0.05 torr /K<br />
PTFE tubing connection 10/8 mm, hose nozzle DN 6/10 mm (KF DN 16 optional)<br />
-10 - 60 / 10 - 40<br />
40 / 80<br />
Coated metal housing<br />
IP 20<br />
IP 54<br />
100-230 V / 50-60 Hz<br />
400 mA (line voltage: 110-230 V) / 340 mA (line voltage: 100-110 V)<br />
7.2(4) A at 250V~ / 7.2(7.2) A at 125 V~<br />
6(4) A at 250 V~ / 6(6) A at 125 V~<br />
163 x 90 x 68<br />
0.8<br />
TM Please order power cable separately<br />
accessories<br />
<strong>Vacuum</strong> controller complete with integrated vacuum<br />
sensor, ready <strong>for</strong> use, with manual.<br />
pg. 167