Chapter 1 Preparing the Instrument<br />
Preparing the Instrument<br />
When to Replenish or Change Polymer<br />
IMPORTANT! Always replace<br />
polymer that has been on the<br />
instrument longer than one week.<br />
Before Using the Polymer<br />
1. Remove the polymer from 4 °C storage.<br />
2. Loosen the cap and bring the polymer to room<br />
temperature.<br />
3. To dissolve deposits, tighten the cap and gently<br />
swirl the polymer.<br />
Replenishing the Polymer<br />
IMPORTANT! Wear gloves while<br />
handling polymer, the capillary array,<br />
septa, or buffer reservoirs.<br />
Notes<br />
Su M T W Th F S<br />
If polymer on the instrument... Then ...<br />
has been on less than one week and is in sufficient quantity<br />
to complete your runs a<br />
has been on less than one week, and insufficient in quantity<br />
to complete your runs<br />
has been on longer than one week<br />
is the wrong type (a change between POP-4, POP-6, and/or<br />
POP-7 polymers is required)<br />
CHEMICAL HAZARD. POP<br />
polymer may cause eye, skin, and respiratory tract<br />
irritation. Please read the MSDS, and follow the<br />
handling instructions. Wear appropriate protective<br />
eyewear, clothing, and gloves. Use for research and<br />
development purposes only.<br />
1. Click < Instrument Name> in the tree pane.<br />
Remove all bubbles, and then proceed with instrument<br />
preparation.<br />
Add fresh polymer to the polymer supply by following the<br />
Replenish Polymer Wizard.<br />
Replace the installed polymer type with a different type by<br />
following the Change Polymer Type Wizard.<br />
a A <strong>3130</strong>xl genetic analyzer run uses 50 to 80 µL of polymer and a <strong>3130</strong> genetic analyzer run uses ~25 to 40µL of polymer.<br />
14 <strong>Applied</strong> Biosystems <strong>3130</strong>/<strong>3130</strong>xl <strong>Genetic</strong> <strong>Analyzers</strong> <strong>Getting</strong> <strong>Started</strong> <strong>Guide</strong><br />