Chapter 1 Preparing the Instrument<br />

Preparing the Instrument<br />

When to Replenish or Change Polymer<br />

IMPORTANT! Always replace<br />

polymer that has been on the<br />

instrument longer than one week.<br />

Before Using the Polymer<br />

1. Remove the polymer from 4 °C storage.<br />

2. Loosen the cap and bring the polymer to room<br />

temperature.<br />

3. To dissolve deposits, tighten the cap and gently<br />

swirl the polymer.<br />

Replenishing the Polymer<br />

IMPORTANT! Wear gloves while<br />

handling polymer, the capillary array,<br />

septa, or buffer reservoirs.<br />

Notes<br />

Su M T W Th F S<br />

If polymer on the instrument... Then ...<br />

has been on less than one week and is in sufficient quantity<br />

to complete your runs a<br />

has been on less than one week, and insufficient in quantity<br />

to complete your runs<br />

has been on longer than one week<br />

is the wrong type (a change between POP-4, POP-6, and/or<br />

POP-7 polymers is required)<br />

CHEMICAL HAZARD. POP<br />

polymer may cause eye, skin, and respiratory tract<br />

irritation. Please read the MSDS, and follow the<br />

handling instructions. Wear appropriate protective<br />

eyewear, clothing, and gloves. Use for research and<br />

development purposes only.<br />

1. Click < Instrument Name> in the tree pane.<br />

Remove all bubbles, and then proceed with instrument<br />

preparation.<br />

Add fresh polymer to the polymer supply by following the<br />

Replenish Polymer Wizard.<br />

Replace the installed polymer type with a different type by<br />

following the Change Polymer Type Wizard.<br />

a A <strong>3130</strong>xl genetic analyzer run uses 50 to 80 µL of polymer and a <strong>3130</strong> genetic analyzer run uses ~25 to 40µL of polymer.<br />

14 <strong>Applied</strong> Biosystems <strong>3130</strong>/<strong>3130</strong>xl <strong>Genetic</strong> <strong>Analyzers</strong> <strong>Getting</strong> <strong>Started</strong> <strong>Guide</strong><br />