ENGLISH<br />
8.4 Characteristics of the water to be used<br />
TABLE SHOWING THE QUALITY LEVELS LAID BY THE UNI EN 13060: 2005<br />
CEN STANDARD UNI EN 13060 : 2005<br />
Evaporation residue ≤ 10 mg/l<br />
Silicon oxide ≤ 1 mg/l<br />
Iron ≤ 0.2 mg/l<br />
Cadmium ≤ 0.005 mg/l<br />
Lead ≤ 0.05 mg/l<br />
mg/l Remains of heavy metals apart<br />
from iron, cardium, lead<br />
≤ 0.1 mg/l<br />
Chloride (Cl’) ≤ 2 mg/l<br />
Phosphate (P20s) ≤ 0.5 mg/l<br />
Conductivity (at 20°C) ≤ 15 µs/cm<br />
Ph value (acidity level) 5 ÷ 7,5<br />
Appearance<br />
Transparent, clear, without<br />
deposits<br />
Hardness (and alkaline earth ions) ≤ 0.02 mmol/l<br />
8.5 Loading the material into the autoclave<br />
Place the materials to be sterilized on the trays provided.<br />
- never superpose the materials<br />
- arrange the wrapped instruments, with the paper part facing upwards<br />
- never bring the materials into contact with the sterilization chamber or the closing door<br />
- put the scissors and dental forceps with the blade open<br />
PICT.23<br />
When you have place all the instruments close the door and you can read<br />
CLOSED on the display.<br />
DOOR<br />
8.6 Starting the sterilization cycle<br />
Choose sterilization program most adaptable of load prepared and press the button<br />
B-SELECT.<br />
Each individual pressure changes program selection to the next.<br />
After having chosen the program start the cycle by pressing the button B-START, the door<br />
will lock automatically and the cycle will begin.<br />
During the cycle, the display will show all parameters and relative information.<br />