Series <strong>320</strong> <strong>XR</strong>Dismantling the weighing cell5. Unscrew 3 screws of the sensor mechanism and gently tilt the sensor mechanism to the rear.6. Remove the two inner assembly pins [9/D3] → unscrew from the balance arm.7. Remove the two outer assembly pins [9/D3] → unscrew from the chassis.8. Remove the two assembly jigs [7/D3] + [8/D3].9. Remove the centring mandril [2/D2], put the weighing cell on side and turn the 3 screw (30/B2)approx. 5mm out, pull the magnet pot carefully down.10.Carefully tilt the balance arm (19/B2) over the magnet and the sensor mechanism and take it out.11.Further dismantling of the weighing cell is unnecessary.<strong>Precisa</strong> Gravimetrics AG Index b2 E8