Tensior T8 - Eijkelkamp Agrisearch Equipment

Tensior T8 - Eijkelkamp Agrisearch Equipment

Tensior T8 - Eijkelkamp Agrisearch Equipment

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

2. Fill the syringe and the connected rubber tube with water. Attach the<br />

syringe to the refilling tube.<br />

3. Continuously inject water into the filling tube while monitoring the pressure<br />

signal at any time. Water and bubbles arise from the de-airing tube. Stop<br />

injecting water when no more air bubbles appear.<br />

16<br />

Caution: Do not let the pressure exceed 2000 hPa !<br />

4. Remove the syringe. Put a drop of water on both rubber and steel tube's<br />

end to avoid that any air will get into the tubes. Connect both.<br />

With this method a measuring range of approx. 800 hPa can be achieved.<br />

5.4 Refilling with vacuum<br />

To reach the full measuring range the Tensiometer water can be degassed<br />

even more completely by using a vacuum. This method can be applied for<br />

installed or not installed Tensiometers, in any installation angle and for refilling<br />

tubes longer than 5 m.<br />

The refilling kit BKTex contains all required tools:<br />

- manual vacuum pump and vacuum bottle with tube and valve<br />

- syringe with valve<br />

Procedure:<br />

1. Connect the <strong>T8</strong> to the measuring device and keep an eye on the pressure<br />

signal at any time. Pull off the rubber tube from the refilling tube.<br />

3, The syringe as well as the tube on the syringe should be filled with degassed<br />

water. Close the valve (!) and connect syringe and refilling tube.<br />

2, Connect vacuum bottle and de-airing tube. Now, evacuate the bottle to the<br />

maximum possible vacuum. This will enlarge the remaining bubble inside the<br />

cup. Briefly open and then close the valve of the syringe to let water flow<br />

into the Tensiometer. The air bubble will be sucked into the vacuum bottle.<br />

Repeat this 2 or 3 times until no bubbles come out anymore.