7 Concept and installation<br />
7.1 Selecting the measuring site<br />
The installation spot should be representative for the soil horizon! Therefore,<br />
in heterogenic soils, classifying drillings should be made before or during<br />
installation.<br />
On farmed sites with vegetation root spreading and root growth during the<br />
measuring period must be considered. Fine roots will grow around the<br />
Tensiometer cup as this is a poor but still secure source of water. Therefore,<br />
avoid the root zone or move the Tensiometer from time to time depending on<br />
the root growth.<br />
Disturbing effects like waysides, the rim of a field, slopes or dints must be<br />
avoided or considered in the interpretation of the measuring results.<br />
7.2 Selecting the installation angle<br />
An installation position would be ideal if the typical water flow is not disturbed<br />
by the Tensiometer. No preferential water flow along the shaft should be<br />
created. Therefore, Tensiometers are preferably installed at an angle.<br />
7.2.1 "Vertical" with downwardly angle<br />
When installed from the surface, an angle of 25° to 65° from the<br />
vertical line is optimal for refilling. In an absolutely vertical position<br />
air bubbles might hide inside the edges of the cup adapter. Still,<br />
they could be removed completely with the vacuum refilling kit<br />
BKTex. In this position, the refilling tube is the shorter stainless<br />
steel tube with the black mark. Into this tube, water is injected for<br />
refilling.<br />
7.2.2 "Horizontal" with upwardly angle<br />
When installed horizontally from inside a well or pit hole,<br />
the Tensiometer must point upwards! This means the<br />
20