10.2 Osmosis<br />
The ceramic used has a pore size of 1 µm. Thus, ions can hardly be kept out.<br />
An influence of the measured value by osmotic effect is negligible under normal<br />
circumstances. If the <strong>T8</strong> is dipped into saturated NaCl solution 10 hPa are<br />
measured. Either consider this fault or fill the water with this solution (after it<br />
has been degassed).<br />
10.3 Emptying the <strong>T8</strong><br />
Freezing of the Tensiometer water will destroy the pressure sensor and<br />
possibly the cup. Therefore, frost endangered Tensiometer must be<br />
removed from the soil or emptied.<br />
In non-freezing depths Tensiometers can remain in the ground in winter.<br />
Procedure for emptying:<br />
1. Remove the connecting rubber tube from the refilling tube.<br />
2. Connect the empty syringe to the refilling tube.<br />
3. Completely suck out the Tensiometer water.<br />
4. Connect rubber tubes and filing tube.<br />
Until now, we received no tensiometer with freezing damage while installed in<br />
the soil deeper than 30 cm, although the soil was frozen.<br />
30