10.2 Osmosis<br />

The ceramic used has a pore size of 1 µm. Thus, ions can hardly be kept out.<br />

An influence of the measured value by osmotic effect is negligible under normal<br />

circumstances. If the <strong>T8</strong> is dipped into saturated NaCl solution 10 hPa are<br />

measured. Either consider this fault or fill the water with this solution (after it<br />

has been degassed).<br />

10.3 Emptying the <strong>T8</strong><br />

Freezing of the Tensiometer water will destroy the pressure sensor and<br />

possibly the cup. Therefore, frost endangered Tensiometer must be<br />

removed from the soil or emptied.<br />

In non-freezing depths Tensiometers can remain in the ground in winter.<br />

Procedure for emptying:<br />

1. Remove the connecting rubber tube from the refilling tube.<br />

2. Connect the empty syringe to the refilling tube.<br />

3. Completely suck out the Tensiometer water.<br />

4. Connect rubber tubes and filing tube.<br />

Until now, we received no tensiometer with freezing damage while installed in<br />

the soil deeper than 30 cm, although the soil was frozen.<br />