7.5 Length of the refilling tubes<br />
Keep the refilling tubes as short as possible. Caused by the thermal expansion<br />
of air, any air inside the tubes will affect the readings when temperature<br />
changes. Preferably, insulate the refilling tubes or install them isothermal into<br />
the soil.<br />
7.6 Jacket tubes<br />
Jacket tubes are useful with shafts longer than 2 m, in pebbly soils or gravel,<br />
and for horizontal installations from inside a well or pit hole. The jacket tube<br />
should end 30 to 50 cm away from the cup so leaking or condensation water is<br />
not conducted to the cup. The inner diameter of the jacket tube should be at<br />
least 35 mm.<br />
7.7 Ideal condition for installation<br />
For the installation of Tensiometers, the ideal conditions are:<br />
- Frost-free soil.<br />
- Coarse clay or loess.<br />
- In soil types loess, coarse clay, loam or clay as well as in highly organic soils<br />
(humus) preferably with a water tension close to saturation (moist).<br />
- In sandy or pebbly soils with a water tension over 100 hPa (rather dry).<br />
- Low skeletal structure (gravel). The more gravel in a soil the more often<br />
the drilling has to be repeated to reach the required depth.<br />
22