21.02.2013 Views

Tensior T8 - Eijkelkamp Agrisearch Equipment

Tensior T8 - Eijkelkamp Agrisearch Equipment

Tensior T8 - Eijkelkamp Agrisearch Equipment

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

User Manual<br />

<strong>T8</strong><br />

Tensiometer<br />

Version January 2003<br />

<strong>T8</strong><br />

P.O. Box 4, 6987 ZG Giesbeek<br />

Nijverheidsstraat 30,<br />

6987 EM Giesbeek,<br />

The Netherlands<br />

T +31 313 880200<br />

F +31 313 880299<br />

E eijkelkamp@eijkelkamp.com<br />

I http://www.eijkelkamp.com


Foreword<br />

The success of any technical system is directly depending on a correct<br />

operation. Moreover, the systems must be reliable, durable and require a<br />

minimum of maintenance to achieve target-directed results and keep the<br />

servicing low.<br />

At the beginning of a measuring task or research project the target, all effective<br />

values and the surrounding conditions must be defined. This leads to the<br />

demands for the scientific and technical project management which describes<br />

all quality related processes and decides on the used methods, the technical<br />

and measurement tools, the verification of the results and the modelling.<br />

The continuously optimised and synergic correlation of all segments and it's<br />

quality assurance are finally decisive for the success of a project.<br />

Thus, I would be pleased to receive your critiques and appraisals.<br />

München, 30.09.2002<br />

3


Table of contents<br />

1 Content of delivery ....................................................................................7<br />

2 Short instructions .......................................................................................7<br />

3 Safety notes................................................................................................9<br />

4 Description of the <strong>T8</strong>...............................................................................10<br />

4<br />

4.1 Output signals.....................................................................................10<br />

4.2 Sensor body........................................................................................10<br />

4.2.1 Pressure sensor.........................................................................10<br />

4.2.2 Temperature sensor .................................................................11<br />

4.2.3 IR indicator................................................................................11<br />

4.3 The ceramic cup .................................................................................12<br />

5 Refilling.....................................................................................................13<br />

5.1 When does the Tensiometer needs to be refilled?.............................13<br />

5.2 Refilling in the lab................................................................................13<br />

5.3 Refilling in the field..............................................................................15<br />

5.4 Refilling with vacuum ..........................................................................16<br />

6 Calibration................................................................................................18<br />

6.1 Checking the offset.............................................................................18<br />

6.2 Offset correction for non horizontal installations...............................19<br />

6.3 Correlation of water column and pressure ........................................19<br />

7 Concept and installation...........................................................................20<br />

7.1 Selecting the measuring site................................................................20<br />

7.2 Selecting the installation angle ............................................................20<br />

7.2.1 "Vertical" with downwardly angle .............................................20<br />

7.2.2 "Horizontally" with upwardly angle ...........................................20


7.3 Number of Tensiometers per level ....................................................21<br />

7.4 Extend of the measuring site ..............................................................21<br />

7.5 Length of the refilling tubes ................................................................22<br />

7.6 Jacket tubes.........................................................................................22<br />

7.7 Ideal condition for installation.............................................................22<br />

7.8 Documentation...................................................................................23<br />

7.9 Installation...........................................................................................23<br />

8 Connecting the <strong>T8</strong> ...................................................................................25<br />

8.1 Single-ended or differential application...............................................25<br />

8.2 Connecting the IR-indicator................................................................26<br />

8.3 Wire specifications..............................................................................26<br />

9 Protecting the measuring site...................................................................27<br />

9.1 Theft and vandalism............................................................................27<br />

9.2 Frost ...................................................................................................27<br />

9.3 Lightning .............................................................................................27<br />

10 Additional notes..................................................................................29<br />

10.1 Interpretation and maximum measuring range..............................29<br />

10.2 Osmosis .........................................................................................30<br />

10.3 Emptying the <strong>T8</strong> ............................................................................30<br />

11 Accessories .........................................................................................31<br />

11.1 Connecting and extension cables...................................................31<br />

11.2 Handheld measuring device ...........................................................31<br />

11.3 Augers............................................................................................31<br />

11.4 Refill- and calibration kits ...............................................................32<br />

12 Contact and communication...............................................................33<br />

13 Technical specifications.......................................................................34<br />

5


1 Content of delivery<br />

The delivery of a <strong>T8</strong> includes:<br />

• Tensiometer, calibrated and filled, cable with plug and protective cap<br />

• This manual<br />

• Plastic bottle protecting the ceramic cup, must be filled to half with<br />

water to keep the cup moist<br />

• Shaft water protection disk<br />

• Calibration certificate with technical data and output signals<br />

Available accessories: see chapter "Accessories"<br />

If you wish to receive a individual calibration certificate for each Tensiometer,<br />

please contact us and give us the serial number of the Tensiometers.<br />

2 Short instructions<br />

Please read and understand the complete manual prior to first use.<br />

In general, the <strong>T8</strong> is filled and ready for installation when supplied.<br />

For installation proceed as follows:<br />

1. Drilling the borehole: An installation angle of 25° to 65° from the vertical<br />

line is ideal for the optimal removal of air from the cup. Mark the required<br />

drilling depth on the auger and on the Tensiometer shaft.<br />

Note: Installation depth = drilling depth / cos α<br />

Drill a borehole, diameter 25 mm, with the required depth. offers<br />

special Tensiometer gouge augers with shaped blade tip.<br />

2. Illuvation of the cup is only needed in clayey soils and only if the borehole is<br />

larger than the diameter of the cup (which is 24 mm). With the special<br />

Tensiometer gouge auger illuvation is unnecessary. In coarse sand and<br />

7


pebbly grounds a fine grained illuvation paste would even effect a water<br />

reservoir which would slow down the response.<br />

3. Take off the protective plastic bottle from the Tensiometer cup. Pull the<br />

bottle it off carefully or slowly turn it counter-clockwise!<br />

4. Insert the <strong>T8</strong> into the borehole up to the depth mark without using force.<br />

In clayey soils an overpressure might develop: Check the Tensiometer's<br />

pressure reading with an Infield measuring device or a datalogger!<br />

8<br />

Important: Pay attention to the engraved black spot on the<br />

end of the shaft that marks the position of the exit opening of<br />

the external filling inside the cup:<br />

a) Downwards installation: If the position of the cup will be<br />

lower than the end of the shaft, this mark must point<br />

upwards!<br />

b) Upwards installation: If the position of the cup<br />

will be higher than the end of the shaft, this mark<br />

must point downwards!<br />

5. Push down the shaft water retaining disk to a position directly on the soil<br />

surface. Water flowing down the shaft will not disturb measurements.<br />

6. Leave on the protective cap on the plug if the plug is not connected.<br />

7. Connect the Tensiometer signal wires as specified (see chapter 8<br />

Connecting the <strong>T8</strong>). If the measuring station is prepared , attach<br />

the Tensiometer plug to the suitable connecting cable.<br />

Notes:<br />

- The less air is inside the cup and the better the soil's conductivity is, the faster<br />

the Tensiometer will respond.<br />

- If the soil is dryer than 900 hPa, it does not make sense to refill the<br />

Tensiometer. Refill the Tensiometer as soon as the Tensiometer that is


installed in the next lower level has reached the reading when the upper<br />

Tensiometer had dried out.<br />

3 Safety notes<br />

Tensiometer are instruments for measuring the soil water tension, soil water<br />

pressure and soil temperature and are designed for this purpose only.<br />

Caution<br />

Lightning: Long cables act as antennas and might conduct surge voltage in<br />

case of lightning stroke – this might damage sensors and instruments.<br />

Caution<br />

Frost: Tensiometers are filled with water and therefore, are sensitive to<br />

frost! In wintertime, do not leave Tensiometers inside your car overnight.<br />

Generally, frost should not harm Tensiometers installed in depths below<br />

30 cm.<br />

Caution<br />

Overpressure: The non destructive maximum pressure is 3000 hPa. Higher<br />

pressure, which might occur for example during insertion in wet clayey<br />

soils or triaxial vessels, will destroy the pressure sensor! The applied<br />

pressure should not exceed 2000 hPa!<br />

Caution<br />

Electronics: Any electrical installations should only be executed by qualified<br />

personnel!<br />

Caution<br />

Ceramic cup: Avoid to touch the cup with your fingers. The ceramic<br />

should not have contact with grease or soap as this will influence the<br />

hydrophilic performance.<br />

9


4 Description of the <strong>T8</strong><br />

4.1 Output signals<br />

The <strong>T8</strong> is optimised for in-the-field use. Signals for soil water tension, soil<br />

temperature and filling status are offered.<br />

Water tension and temperature are linear voltage signals 0 ... 2 V. Thus, the<br />

signals are directly connectable to nearly any datalogger or other data<br />

recording device.<br />

The IR indicator works like a switch so the filling status can have two levels:<br />

"0V" (switch is open) for correct filling, or Vin = power supply voltage (switch<br />

is closed) if the Tensiometer needs to be refilled.<br />

4.2 Sensor body<br />

The complete electronic of the <strong>T8</strong> is integrated in the sensor body. The<br />

housing consists of glass-fibre reinforced synthetic. All electronic parts are<br />

completely covered with epoxy resin - the optimal and everlasting protection<br />

against moisture.<br />

The <strong>T8</strong> is calibrated electronically by<br />

programmable potentiometers. The calibration is<br />

stored in EPROM's.<br />

4.2.1 Pressure sensor<br />

The piezoelectric pressure sensor measures the soil water tension against the<br />

atmospheric pressure. The atmospheric pressure is conducted through a<br />

watertight diaphragm (the white, 2 cm long tube on the cable) and through the<br />

cable to the reference side of the pressure sensor.<br />

The non destructive maximum pressure is 3.000 hPa. Higher pressure will<br />

damage the sensor and absolutely must be avoided!<br />

10<br />

Sensors and refilling tube on sensor body


4.2.2 Temperature sensor<br />

A Pt1000 1/3 DIN B is used as temperature sensor. The Pt1000 has a<br />

maximum deviation of ±0.1 K at 0°C.<br />

The tip of the sensor dips into the Tensiometer cup's water. Thus, the best<br />

possible thermal contact to the soil is achieved.<br />

The correlation of water tension/water content is temperature depended. The<br />

influence is low with water tensions from 0 … 100 hPa ⇒ 0 … 6 hPa/K, but<br />

with tensions over 1000 hPa it is:<br />

Ψ = (R T/M) ln p/po Ψ = water tension R = gas constant (8,31J/mol K)<br />

M = molecular weight p = vapour pressure<br />

po = saturation vapour pressure at soil temperature<br />

(see vapour pressure tables by Scheffler/Straub, Grigull)<br />

4.2.3 IR indicator<br />

With the IR indicator the filling status of a <strong>T8</strong> in a downwardly installation can<br />

be checked without removing the Tensiometer from the soil.<br />

If the Tensiometer is installed in an upwardly angle (cup is higher than the<br />

end of the shaft), the indicator will react not before the cup is almost<br />

completely empty.<br />

Measuring principle of the IR indicator: A photo-sensor and a source of IR-light<br />

are integrated in the <strong>T8</strong> body. If there is only water between source and sensor<br />

the absorption is high and the photo-sensor will not respond. If there should be<br />

a bubble in-between, the measured light intensity arises and the electronic will<br />

switch to the high output signal, indicating the need to refill. The IR-"switch"<br />

can be burdened with 50 mA. LEDs or relays can be connected.<br />

Benefits of the IR indicator are an optimal service: If for example mounting rails<br />

with integrated LEDs are used, the user can check at a glance if all<br />

Tensiometers work well or which need to be refilled.<br />

11


Also, for quality assurance, the status of the signal can be logged for later<br />

evaluation of the readings.<br />

4.3 The ceramic cup<br />

Avoid to touch the cup with your fingers. The ceramic should not have<br />

contact with grease or soap as this will influence the hydrophilic<br />

performance.<br />

To transfer the soil water tension as a negative pressure into the Tensiometer,<br />

a semi-permeable diaphragm is required. This must have good mechanical<br />

stability and water-permeability, but also have gas impermeability.<br />

The <strong>T8</strong> cup consists of ceramic Al2O3 sinter material. The special<br />

manufacturing process guarantees homogeneous porosity with good water<br />

conductivity and very high firmness. Compared to conventional porous ceramic<br />

the cup is much more durable.<br />

The bubble point is higher than 6000 hPa. If the soil is dryer than 6000 hPa the<br />

negative pressure inside the cup decreases and the readings go down to 0 hPa.<br />

With these characteristics this material has outstanding suitability to work as<br />

the semi permeable diaphragm for Tensiometers.<br />

4.4 Reference air pressure<br />

The reference atmospheric air pressure is conducted to the pressure<br />

transducer via the air permeable (white) Teflon membrane and through the<br />

cable. The membrane does not absorb water. Water will not pass through the<br />

membrane into the cable, but condensed water inside the cable will leave the<br />

cable through the membrane.<br />

The white membrane on the cable must always have contact to air and<br />

should never be submersed into water.<br />

12


5 Refilling<br />

To assure a rapid and reliable measurement of the soil water tension, the cup<br />

must possibly be filled with de-ionised and degassed water and without bubbles<br />

of air. After dry periods or periods with a large number of wet and drying out<br />

successions, the <strong>T8</strong> must be refilled.<br />

For all refilling methods, the following items are always required:<br />

- Syringe with valve<br />

- Degassed, de-ionised water<br />

- Measuring device for Tensiometer pressure signal<br />

5.1 When does the Tensiometer needs to be refilled?<br />

The Tensiometer needs to be refilled:<br />

a, For Tensiometers installed "downwardly", the IR indicator will light up.<br />

b, If no signal peaks show up anymore.<br />

c, Tensiometers should not be refilled until the soil is moister than 900 hPa.<br />

If the soil gets dryer than 805 hPa, the readings will stop at the vapour pressure<br />

of water (i. e. 927 hPa at 20°C and atmospheric pressure of 950 hPa). By<br />

diffusion and slight leakage the reading will slowly drop within months.<br />

Still, if the soil dries out more than 6000 hPa, the negative pressure will drop<br />

much faster as air will enter the cup.<br />

5.2 Refilling in the lab<br />

To reach a measuring range of 900 hPa the Tensiometers should be refilled in<br />

the laboratory. With this refilling method a measuring range up to approx. 900<br />

hPa can be achieved (at 1013 hPa atmospheric pressure). A refill kit is required.<br />

Degassing procedure:<br />

1, Set up the refilling kit and connect the vacuum pump. The pump should<br />

achieve at least 30 hPa against vacuum (approx. 930 hPa).<br />

13


2, Unscrew the cup in clockwise direction and empty it. Do not touch the<br />

ceramic cup with your fingers. Wrap a clean towel around the cup.<br />

The pressure sensor diaphragm is inside the small boring on the pressure<br />

sensor body. It is very sensitive and may never be touched! It can be<br />

destroyed even by slightest contact! No contamination should get on the<br />

sealing and gasket.<br />

3, If the cup is drained it should be drenched in distilled water overnight by<br />

placing it in a beaker filled with water.<br />

Important: There should be no water inside the cup. The cup must be in an<br />

upright position so only the cup's outside is covered with water. Otherwise<br />

cavities of air might be enclosed inside the ceramic.<br />

4, The drenched cup should be empty. Insert the cup to the adapter and<br />

connect it to the degassing device. Place the cup in water in an upright<br />

position as described above.<br />

5, Fill the other adapter capsule to the half with water and insert the<br />

sensor body. Connect the adapter to the degassing device as well.<br />

6, Start the vacuum pump. With drenched cups, the procedure will take 4 to 5<br />

hours. From time to time, tap on cup and sensor body to loosen bubbles.<br />

Degassing is complete when no air bubbles ascend from ceramic and body<br />

and the cup is completely filled with water.<br />

14<br />

Beaker<br />

Ceramic<br />

cup<br />

Sensor<br />

body<br />

Vacuum<br />

bottle<br />

Pressure reducer<br />

Vacuum pump


7, Before screwing the cup onto the body, connect the sensor to a measuring<br />

device to observe the pressure signal. Carefully and slowly screw the cup on<br />

the sensor body. Make sure that no bubbles are enclosed.<br />

Caution: The pressure must not exceed 2000 hPa!<br />

8, Now, the cup and the pressure sensor are completely degassed. Finally, the<br />

refilling tubes have to be filled as well. To do so, continue with the<br />

procedure as described in chapter 5.3 or 5.4.<br />

5.3 Refilling in the field<br />

For an easy refilling of still saturated cups on site and without removing the<br />

Tensiometer from the soil, the Tensiometers can be refilled externally through<br />

the high-grade steel capillary tubes. If the refilling tubes are longer than 5 m a<br />

vacuum pump might be necessary – see chapter 5.4.<br />

Procedure:<br />

1. Connect the <strong>T8</strong> to the measuring device and keep an eye on the pressure<br />

signal at any time.<br />

2. Two steel capillary tubes come out from the <strong>T8</strong> shaft: the refilling tube and<br />

the de-airing tube. Pull off the rubber tube from the refilling tube.<br />

Left fig: Downwardly installation – the<br />

marked tube is the refilling tube<br />

Right fig: Upwardly installation – the<br />

marked tube is the de-airing tube<br />

15


2. Fill the syringe and the connected rubber tube with water. Attach the<br />

syringe to the refilling tube.<br />

3. Continuously inject water into the filling tube while monitoring the pressure<br />

signal at any time. Water and bubbles arise from the de-airing tube. Stop<br />

injecting water when no more air bubbles appear.<br />

16<br />

Caution: Do not let the pressure exceed 2000 hPa !<br />

4. Remove the syringe. Put a drop of water on both rubber and steel tube's<br />

end to avoid that any air will get into the tubes. Connect both.<br />

With this method a measuring range of approx. 800 hPa can be achieved.<br />

5.4 Refilling with vacuum<br />

To reach the full measuring range the Tensiometer water can be degassed<br />

even more completely by using a vacuum. This method can be applied for<br />

installed or not installed Tensiometers, in any installation angle and for refilling<br />

tubes longer than 5 m.<br />

The refilling kit BKTex contains all required tools:<br />

- manual vacuum pump and vacuum bottle with tube and valve<br />

- syringe with valve<br />

Procedure:<br />

1. Connect the <strong>T8</strong> to the measuring device and keep an eye on the pressure<br />

signal at any time. Pull off the rubber tube from the refilling tube.<br />

3, The syringe as well as the tube on the syringe should be filled with degassed<br />

water. Close the valve (!) and connect syringe and refilling tube.<br />

2, Connect vacuum bottle and de-airing tube. Now, evacuate the bottle to the<br />

maximum possible vacuum. This will enlarge the remaining bubble inside the<br />

cup. Briefly open and then close the valve of the syringe to let water flow<br />

into the Tensiometer. The air bubble will be sucked into the vacuum bottle.<br />

Repeat this 2 or 3 times until no bubbles come out anymore.


3, Close the valve of the vacuum tube and remove it. Now press down the<br />

syringe to fill in another amount of approx. 5 ml of water. Remove the<br />

syringe and connect both steel tubes with the rubber connecting tube.<br />

Fig: Downwardly installation – the marked tube is the refilling tube<br />

Fig: Upwardly installation – the marked tube is the de-airing tube<br />

17


6 Calibration<br />

Programmable potentiometers are integrated into the <strong>T8</strong>. They can be<br />

addressed via the two digital lines.<br />

Although the electronics is completely covered with epoxy resin, it is still<br />

possible to calibrate the sensor.<br />

The following parameters can be calibrated:<br />

• Pressure sensor: inclination and offset<br />

• Temperature sensor: offset<br />

• IR indicator: trigger level<br />

The <strong>T8</strong> Programming Unit KB<strong>T8</strong> is required to address the potentiometers<br />

and to check the inclination.<br />

6.1 Checking the offset<br />

There a two possible ways to check the offset:<br />

1. Put the filled <strong>T8</strong> into a beaker with water and immerse the cup for approx.<br />

3 to 4 cm. Wait for the value to stabilise. The measured signal must be<br />

between -5 hPa and +5 hPa (at room temperature). If air bubbles are inside<br />

the Tensiometer, the reading might take a while to reach this value.<br />

2. A more reliable method would be to screw of the cup. Then, shake the<br />

body to remove the water inside the pressure sensor. The pressure value<br />

must be in the range of -5 hPa to +5 hPa. Before screwing the cup on the<br />

<strong>T8</strong>-body, pressure sensor and cup must be refilled with degassed and deionised<br />

water (see chapter "Filling in the laboratory").<br />

Caution: The pressure sensor diaphragm inside the small boring on the<br />

pressure sensor body is very sensitive and may never be touched! It can be<br />

destroyed even by slightest contact!<br />

Also, no contamination should get on the sealing and gasket.<br />

18


6.2 Offset correction for non horizontal installations<br />

The <strong>T8</strong> is calibrated for horizontal installation. If it is built in a non horizontal<br />

position, the vertical water column drawing on the pressure sensor has to be<br />

compensated. With vertical installation (completely upright with 0° angle) the<br />

deviation is largest. In this case a water column of 5 cm additionally draws on<br />

the pressure sensor and causes a shift of -5 hPa. This means that a value of<br />

-5 hPa will be indicated instead of the actual value of 0 hPa.<br />

Find the correction value for other angles in the table:<br />

Angle to vertical line 0° 10° 15° 20° 25° 30°<br />

Offset correction in [hPa] = [mV] -5 -4,9 -4,8 -4,7 -4,5 -4.3<br />

Angle to vertical line 45° 60° 70° 75° 80° 90°<br />

Offset correction in [hPa] = [mV] -3,5 -2.5 -1,7 -1,3 -0,9 0<br />

6.3 Correlation of water column and pressure<br />

A precision pressure gauge or a water column is needed. Please also observe<br />

the correlation of pressure and water column if the Tensiometer is used as a<br />

piezometer or water level transmitter.<br />

Pressure = density x gravity x height<br />

Density of water at 20°C: 0,998205 kg/cubic decimetres; at 4°C: 1,0 kg/dm3 .<br />

[Pa] = [N]/[qm] [N] = [kg/m qsec]<br />

[kg] = [Pa qm/m] [Pa] = kg/[qqm qsec]<br />

A water column of 1 meter causes a pressure p of:<br />

p [Pa= N/qm] = 998,205 kg/qqm x 9,81 m/qs x 1m<br />

p = 9792,39 [N/qm x Pa qs/m /qqm x m/qs x m] = 97,92 hPa<br />

In reverse a pressure of 100 hPa at 20°C is caused by a water column of<br />

102,15 cm.<br />

19


7 Concept and installation<br />

7.1 Selecting the measuring site<br />

The installation spot should be representative for the soil horizon! Therefore,<br />

in heterogenic soils, classifying drillings should be made before or during<br />

installation.<br />

On farmed sites with vegetation root spreading and root growth during the<br />

measuring period must be considered. Fine roots will grow around the<br />

Tensiometer cup as this is a poor but still secure source of water. Therefore,<br />

avoid the root zone or move the Tensiometer from time to time depending on<br />

the root growth.<br />

Disturbing effects like waysides, the rim of a field, slopes or dints must be<br />

avoided or considered in the interpretation of the measuring results.<br />

7.2 Selecting the installation angle<br />

An installation position would be ideal if the typical water flow is not disturbed<br />

by the Tensiometer. No preferential water flow along the shaft should be<br />

created. Therefore, Tensiometers are preferably installed at an angle.<br />

7.2.1 "Vertical" with downwardly angle<br />

When installed from the surface, an angle of 25° to 65° from the<br />

vertical line is optimal for refilling. In an absolutely vertical position<br />

air bubbles might hide inside the edges of the cup adapter. Still,<br />

they could be removed completely with the vacuum refilling kit<br />

BKTex. In this position, the refilling tube is the shorter stainless<br />

steel tube with the black mark. Into this tube, water is injected for<br />

refilling.<br />

7.2.2 "Horizontal" with upwardly angle<br />

When installed horizontally from inside a well or pit hole,<br />

the Tensiometer must point upwards! This means the<br />

20


cup is in a higher position than the end of the shaft. An upward angle of approx.<br />

5° is ideal for refilling. Note that now de-airing and refilling tube are switched:<br />

the refilling tube is the longer stainless steel tube without the black mark. Into<br />

this tube, water is injected for refilling.<br />

The IR-indicator will react not until at least a half of the cup is empty.<br />

7.3 Number of Tensiometers per level<br />

To verify the results, at least three Tensiometers per soil level and with the<br />

identical soil, vegetation and climatic conditions should be implemented. Thus,<br />

irregular values as well as the variableness of the water situation can be<br />

determined. Guiding principle: More heterogeneous sites and soil structures<br />

require a higher number of Tensiometers.<br />

7.4 Extend of the measuring site<br />

Large distance along with high equidistance between the measuring spots will<br />

reduce the influence of sectional heterogeneity. Therefore, equidistant spots<br />

are mostly used to measure the heterogeneity of the soil, the cultivation or the<br />

observed development.<br />

The following points limit the maximum length for the signal cables:<br />

- Required accuracy: In single-ended applications every 20 m of cable reduces<br />

the accuracy for about 2 hPa or 0,1°C. (Please read chapter "Connecting<br />

the <strong>T8</strong>" on how to compensate the voltage drop of sensors which are<br />

connected single-ended.)<br />

- Local danger of lightning stroke: Cables act as antennas and therefore<br />

should be as short as possible.<br />

Lengths of more than 100 m are possible but not recommendable. The<br />

available connecting cables are described in the chapter "Accessories".<br />

21


7.5 Length of the refilling tubes<br />

Keep the refilling tubes as short as possible. Caused by the thermal expansion<br />

of air, any air inside the tubes will affect the readings when temperature<br />

changes. Preferably, insulate the refilling tubes or install them isothermal into<br />

the soil.<br />

7.6 Jacket tubes<br />

Jacket tubes are useful with shafts longer than 2 m, in pebbly soils or gravel,<br />

and for horizontal installations from inside a well or pit hole. The jacket tube<br />

should end 30 to 50 cm away from the cup so leaking or condensation water is<br />

not conducted to the cup. The inner diameter of the jacket tube should be at<br />

least 35 mm.<br />

7.7 Ideal condition for installation<br />

For the installation of Tensiometers, the ideal conditions are:<br />

- Frost-free soil.<br />

- Coarse clay or loess.<br />

- In soil types loess, coarse clay, loam or clay as well as in highly organic soils<br />

(humus) preferably with a water tension close to saturation (moist).<br />

- In sandy or pebbly soils with a water tension over 100 hPa (rather dry).<br />

- Low skeletal structure (gravel). The more gravel in a soil the more often<br />

the drilling has to be repeated to reach the required depth.<br />

22


7.8 Documentation<br />

For every measuring spot you should:<br />

- Measure out the position where the pressure sensor will be placed. (A must<br />

for installations below the ground surface).<br />

- Take documenting photos before, during and after installation.<br />

- Save a soil sample.<br />

- Write down installation depth and angle with each sensor identification<br />

(serial number).<br />

- Mark all connecting cables with the corresponding sensor identification on<br />

each end.<br />

7.9 Installation<br />

For the installation of the <strong>T8</strong> in the field the following tools are required:<br />

• Tensiometer auger with diameter 25 mm, ideally the<br />

Tensiometer gouge auger with shaped blade tip.<br />

• Rule, spirit level, goniometer, marker pen<br />

• Minute book, maybe camera for documentation of site and soil profile<br />

• Perhaps PE-plastic bags for taking soil samples from the site<br />

Installation procedure:<br />

1. Measure and mark the required drilling depth on auger and Tensiometer<br />

shaft. Drill a hole with the desired depth on the chosen measuring spot.<br />

Read the chapter "Selecting the installation position" for the best installation<br />

angle. Be cautious with the last 20 cm and take out and save this soil.<br />

2. For soils with poor capillary contact or when using augers with a diameter of<br />

over 25 mm, mix a paste of water and crumbled soil material taken out of<br />

the borehole. Fill the paste into the bottom area of the borehole using a<br />

tube (diameter 20 mm).<br />

3. Take off the protective plastic cap from the Tensiometer cup.<br />

23


Important: Only turn of the cap counter-clockwise! Do not store the<br />

Tensiometer without the protective plastic bottle since the cup drains fast!<br />

The bottle must be filled with some water for storage!<br />

4. Connect the Tensiometer to a readout unit. Carefully insert the <strong>T8</strong> into the<br />

borehole up to the stop.<br />

24<br />

Do not use any force. The Tensiometer may not be pounded, since cup or<br />

pressure sensor can be destroyed.<br />

Especially in clayey soils the pressure reading must be monitored as high<br />

pressures might build up! The maximum pressure must not exceed 2000<br />

hPa<br />

5. Push the shaft water retaining disk down to cover the soil surface. This<br />

prevents water from running into the borehole and down along the shaft.<br />

6. Keep the protective plastic cap on the plug whenever the plug is not<br />

connected!<br />

7. Connect the signal cables as described in the chapter "Connecting the <strong>T8</strong>".<br />

8. The Tensiometer will respond to changes in the soil water tension faster if<br />

there is no air inside the system and the soil water conductivity/mobility is<br />

high.<br />

9. Write down serial number, position, installation angle and depth.


8 Connecting the <strong>T8</strong><br />

The <strong>T8</strong> is fitted with an 8-pin plug. The plug can be connected to a Infield 7<br />

handheld measuring device. The Infield 7 displays and stores the soil water<br />

tension, the soil temperature and the filling status.<br />

With the extension cables offered as accessories the <strong>T8</strong> can be<br />

connected to a Datalogger or any other data acquisition devices.<br />

Please observe the max. supply voltage of 20 VDC. Never connect the signal<br />

outputs of the <strong>T8</strong> to a supply voltage source!<br />

The Tensiometer plug must be covered with the added protective cap<br />

anytime the <strong>T8</strong> is not connected to an extension cable.<br />

8.1 Single-ended or differential application<br />

In general a datalogger with differential inputs should be preferred. This means<br />

both the plus signal as well as the minus signal of each sensor are measured. On<br />

many dataloggers less channels are required for single-ended measurements,<br />

which means only the plus signals are measured and related to the common<br />

sensor supply ground. In this case, the voltage drop, depending on the current<br />

consumption and the cable resistance, influences the reading and must be<br />

compensated:<br />

Current consumption: 7 mA; cable resistance: 82 Ω/km<br />

⇒ Voltage drop inside a cable with length 10 m:<br />

UError = R x I = 0,82 Ω x 0,007 A = 5,66 mV<br />

⇒ Error for water tension (1 mV equal to 1 hPa): 5,66 mV = 5,66 hPa<br />

⇒ Error for temperature (1°C equal to 20 mV): 5,66 mV = 0,283 °C<br />

Therefore, the characteristic lines are lifted for +5,66 hPa resp. +0,283 °C per<br />

10 m cable. The incline is not influenced. For each 10 meters of cable,<br />

withdraw 5,66 hPa for pressure (Note: the water tension signal is regarded<br />

positive) and 0,283 °C for temperature from the measured signal value to get<br />

the correct reading.<br />

25


8.2 Connecting the IR-indicator<br />

Some datalogger types require a pull-down resistor (for example 10 kOhm)<br />

against ground for the IR-indicator. Other logger types already have an internal<br />

pull-down resistor, or a pull-up resistor to 5 V, for example for connecting a<br />

reed switch. The external pull-down resistor must be smaller than the logger<br />

internal pull-up resistor or at least be in a relation so the logger will recognize<br />

the zero level.<br />

8.3 Wire specifications<br />

26<br />

white = V in Power supply plus, 6 … 20 V DC<br />

brown = GND Power supply minus<br />

green = Signal + Soil water tension plus<br />

red = Signal Temp Temperature plus<br />

yellow = Signal - Minus for soil water tension and temperature<br />

grey = Signal IR IR-indicator, switched with Vin pink = Calibration SDA (digital)<br />

blue = Calibration SCL (digital)


9 Protecting the measuring site<br />

9.1 Theft and vandalism<br />

The site should be protected against theft and vandalism as well as against any<br />

farming or field work. Therefore, the site should be fenced and signposts could<br />

give information about the purpose of the site. Cables should be protected<br />

against rodents.<br />

9.2 Frost<br />

Tensiometer are filled with water and therefore are endangered by frost!<br />

With temperatures below–5°C do not leave filled Tensiometers in your car,<br />

in a measuring hut, etc.<br />

Do not fill the Tensiometers with Ethanol, as this is corrosive for some<br />

materials (i. e. PMMA) and might destroy these..<br />

Also it is not recommendable to fill the Tensiometers with Decalin, monoethylene-glycol,<br />

di-ethylene-glycol, etc. These could harm any of the materials,<br />

destroy the ceramic cup or leak into the soil.<br />

Specially designed Tensiometers which can be filled with Ethanol are available<br />

on request. They can be used to temperatures down to - 20°C.<br />

9.3 Lightning<br />

Measuring instruments in the field are always endangered by over-voltages.<br />

Whenever technically feasible, over-voltage and false polarity protection are<br />

realised. If you have questions about an optimal integration of the <strong>T8</strong> into a<br />

measuring system, please contact our system engineers.<br />

Still, there is no absolute certain lightning protection! Lightning strokes are not<br />

predictable. The more extensive a site is, the more import is the installation of<br />

a protection system which could be passive with ground spears – preferably<br />

27


with groundwater contact, or active with each sensor and logger individually<br />

equipped with a lightning protection module. This normally is quite costly.<br />

28


10 Additional notes<br />

10.1 Interpretation and maximum measuring range<br />

Since no pressure smaller than vacuum can appear in the Tensiometer, the<br />

atmospheric pressure (reference pressure) limits the measuring range of the<br />

Tensiometers. Additionally the measuring range is reduced by the vapour<br />

pressure of the Tensiometer's filling water, depending on the temperature.<br />

At 20 °C the vapour pressure is 23.4 hPa. The maximum measuring range at an<br />

atmospheric pressure of 950 hPa is calculated as:<br />

950 hPa - 23.4 hPa = 926.6 hPa<br />

The vapour pressure exponentially rises with temperature. A higher ambient<br />

temperature and lower atmospheric pressure reduces the measuring range.<br />

The above considerations are ruled out by the boiling point shift, enabling a<br />

Tensiometer to measure beyond the vapour pressure. This status is<br />

reproducible, but cannot be quantified.<br />

If the soil gets dryer than 850 hPa the reading will halt at the value of the steam<br />

point (i. e. 927 hPa at 20°C and 950 hPa atmospheric pressure) By diffusion and<br />

minor leakage this value will drop within months.<br />

If the soil gets dryer than 10000 hPa the reading will drop much faster as air<br />

will enter the cup. Evaluating this will give further information, especially if<br />

measurements are taken with TDR probes at the same time, as with this the<br />

extrapolation of the pF-curve in the range of 850 … 10000 hPa determined.<br />

Vapour pressure influence on water tension/water content relation:<br />

If temperature rises from 20°C to 25°C in a soil sample with constant water<br />

content, the water tension gets smaller about 8.5 hPa.<br />

Temperature 4°C 10°C 16°C 20°C 25°C 30°C 50°C 70°C<br />

Change of pressure per 0,6 0,9 1,2 1,5 1,9 2,5 7,2 14<br />

Kelvin in [hPa]<br />

29


10.2 Osmosis<br />

The ceramic used has a pore size of 1 µm. Thus, ions can hardly be kept out.<br />

An influence of the measured value by osmotic effect is negligible under normal<br />

circumstances. If the <strong>T8</strong> is dipped into saturated NaCl solution 10 hPa are<br />

measured. Either consider this fault or fill the water with this solution (after it<br />

has been degassed).<br />

10.3 Emptying the <strong>T8</strong><br />

Freezing of the Tensiometer water will destroy the pressure sensor and<br />

possibly the cup. Therefore, frost endangered Tensiometer must be<br />

removed from the soil or emptied.<br />

In non-freezing depths Tensiometers can remain in the ground in winter.<br />

Procedure for emptying:<br />

1. Remove the connecting rubber tube from the refilling tube.<br />

2. Connect the empty syringe to the refilling tube.<br />

3. Completely suck out the Tensiometer water.<br />

4. Connect rubber tubes and filing tube.<br />

Until now, we received no tensiometer with freezing damage while installed in<br />

the soil deeper than 30 cm, although the soil was frozen.<br />

30


11 Accessories<br />

The following equipment is available:<br />

11.1 Connecting and extension cables<br />

Connecting cables for connections to dataloggers, etc.. One end with female<br />

plug M12/IP67 for connection to a Tensiometer, the other end with loose<br />

wires:<br />

article: CC-8/5 (length 5 m)<br />

CC-8/10 (length 10 m)<br />

CC-8/20 (length 20 m)<br />

Extension cables with one each male and female plug M12/IP67:<br />

article: EC-8/10 (length 10 m)<br />

EC-8/20 (length 20 m)<br />

Individual lengths on request.<br />

11.2 Handheld measuring device<br />

Infield 7 handheld measuring device for manual readings (with or<br />

without calibration function). article: INFIELD7.<br />

11.3 Augers<br />

Tensiometer gouge auger with specially shaped blade. The tip of the blade has<br />

the same shape and diameter as the Tensiometer cup, so no illuvation with clay<br />

paste is necessary. article: TB-25<br />

31


11.4 Refill and calibration kits<br />

32<br />

BKTex<br />

Refill kit for external refillable Tensiometer, including<br />

manual vacuum pump, 250 ml coated glass bottle,<br />

refill syringe, pipes and valves. article: BKTex<br />

BKT468<br />

Refill kit for Tensiometer, incl. base stand with<br />

clamps, adapter that fits to T4, T6 and <strong>T8</strong>-sensor<br />

bodies, 500 ml bottle, pressure gauge, tubes, beaker,<br />

refilling syringe. article: BKT468<br />

KB<strong>T8</strong><br />

Calibration kit for <strong>T8</strong> with calibrator for digital balance, offset, incline; incl.,<br />

precision manometer, valve, mounting kit, vacuum tube, requires a precision<br />

voltmeter. article: KB<strong>T8</strong><br />

All glass vacuum bottles are plastic covered and implosion proof.


12 Contact and communication<br />

Please take advantage of our consulting service – we want you to achieve<br />

optimal results with our instruments.<br />

With pleasure we would like to receive your comments, inspirations, critics.<br />

P.O. Box 4, 6987 ZG Giesbeek<br />

Nijverheidsstraat 30,<br />

6987 EM Giesbeek,<br />

The Netherlands<br />

T +31 313 880200<br />

F +31 313 632167<br />

E eijkelkamp@eijkelkamp.com<br />

I http://www.eijkelkamp.com<br />

33


13 Technical specifications<br />

Dimension, material<br />

Ceramic<br />

Length 60 mm, diameter 24 mm<br />

Sensor body<br />

Material PA6<br />

Shaft<br />

Material PMMA, diameter 25 mm<br />

Cable<br />

Standard cable length<br />

with shafts < 120 cm 1,5 m from sensor body<br />

with shaft > 121 cm 0,6 m from end of shaft<br />

Plug 8-pin, thread M12,<br />

protective rate IP67 (waterproof)<br />

Measuring range<br />

Soil water tension<br />

electronically -1000 hPa ... +1000 hPa<br />

physically<br />

-1000 hPa ... +850 hPa<br />

Temperature -30 ... +70°C<br />

Signal<br />

Soil water tension 0 ... 2 V = -1000 hPa ... +1000 hPa<br />

Temperature<br />

0 ... 2 V = -30 ... +70°C<br />

IR-indicator<br />

Switch open: refilling ok<br />

Switch closed (Vin): refilling required;<br />

max. burden 50mA<br />

Accuracy<br />

Soil water tension ± 5 hPa<br />

Temperature<br />

± 0,2 K (-10...+30°C); ± 0,4 K<br />

Supply<br />

Power supply 6 ... 20 V<br />

Current consumption 7 mA<br />

34

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!