Scale-up<br />

During the optimization process, tests<br />

should be run at three separate scales,<br />

and Millistak+ HC filters are available<br />

in device sizes to match these steps.<br />

Initial experiments can be run on<br />

the two smaller Millistak+ HC filters.<br />

Care should be taken to operate<br />

in a constant-flux manner, recording<br />

the pressure increase. If desired,<br />

<strong>Millipore</strong>’s Pmax SM system can be<br />

employed, to increase the efficiency<br />

of interpreting the test data.<br />

First, to determine the correct type<br />

of Millistak+ HC filters to incorporate<br />

into the process, screening with test<br />

volumes of 100–1000 mL can be<br />

performed with the Millistak+ HC Mini<br />

Capsule (23 cm 2 ). To normalize for<br />

variability in filter media and process<br />

stream, multiple tests (at least three)<br />

are recommended, with different<br />

batches of test fluid.<br />

The next step would be to process<br />

5–25 L of solution through an Opticap<br />

capsule filter with Millistak+ HC media<br />

(650 cm 2 ). Again, testing multiple<br />

devices with multiple process samples<br />

will ensure that the sizing is as accurate<br />

as possible.<br />

When the process has been well<br />

characterized, the process-scale<br />

device can be tested, on volumes<br />

greater than 50 L. Each processscale<br />

16-inch 8-cell cartridge<br />

provides 1.8 m 2 of filtration area.<br />

Care should be taken to<br />

simulate as closely as possible the<br />

final process conditions. Since biological<br />

process streams are very complex<br />

mixtures, their properties could change<br />

drastically if any process parameters<br />

have been changed. If the temperature<br />

of the fluid is not kept constant,<br />

if the age of the fermentation is<br />

changed or if it is stored overnight<br />

under refrigeration, the solution may<br />

change—precipitants may form,<br />

proteins may denature, etc.—and<br />

the sample will, consequently, no<br />

longer be representative of the final<br />

process fluid.<br />

Millistak+ HC Filters Application Guide 19