Flushing<br />
Millistak+ HC filters require a 100 L/m 2<br />
water flush (RO or WFI) prior to use.<br />
The water used for flushing should be<br />
sent directly to drain, since there may<br />
be a small quantity of particles released<br />
during the flush.<br />
To ensure that the media is fully<br />
wetted, a flux of 1 Lpm/ft 2 (645 LMH)<br />
is recommended. To flush a full 8-cell<br />
stack, use the following procedure:<br />
1. Open the vent, to purge any air<br />
from inside the housing.<br />
2. Fill the housing at a low flow rate,<br />
equivalent to about 0.25 Lpm/ft2 (162 LMH).<br />
3. Once the housing is filled, close<br />
the vent. Flow will now be sent<br />
through the filter.<br />
4. Over about 30 seconds, gradually<br />
increase the flow, to achieve a flux<br />
of 1 Lpm/ft2 (645 LMH).<br />
5. Once the full flux (645 LMH) has<br />
been reached, start the flush.<br />
Millistak+ HC Filters Application Guide<br />
Leak-Testing Installed<br />
Cartridge Assemblies<br />
To check that an assembled stack<br />
has been put together correctly,<br />
use the following procedure:<br />
1. Wet out the filters as described<br />
above.<br />
2. Drain the housing. (This may be<br />
done with a connection to drain<br />
through the inlet to the housing.)<br />
3. Once the housing is empty, apply<br />
a forward pressure of 1 – 2 psid<br />
of air pressure.<br />
4. Wait 3 – 5 minutes for stabilization,<br />
since the air pressure will force<br />
some of the retained water within<br />
the media through the stack.<br />
5. Measure the flow of air into the<br />
housing.<br />
There will be some air flow due to the<br />
low bubble point of the media, and<br />
hence some air permeation through<br />
the media will occur. A typical value<br />
of air flow through a fully-wetted<br />
8-cell stack (1.8 m2 ), with a pressure<br />
differential of 2.0 psid, is 100 to<br />
400 cc/min.<br />