Flushing<br />

Millistak+ HC filters require a 100 L/m 2<br />

water flush (RO or WFI) prior to use.<br />

The water used for flushing should be<br />

sent directly to drain, since there may<br />

be a small quantity of particles released<br />

during the flush.<br />

To ensure that the media is fully<br />

wetted, a flux of 1 Lpm/ft 2 (645 LMH)<br />

is recommended. To flush a full 8-cell<br />

stack, use the following procedure:<br />

1. Open the vent, to purge any air<br />

from inside the housing.<br />

2. Fill the housing at a low flow rate,<br />

equivalent to about 0.25 Lpm/ft2 (162 LMH).<br />

3. Once the housing is filled, close<br />

the vent. Flow will now be sent<br />

through the filter.<br />

4. Over about 30 seconds, gradually<br />

increase the flow, to achieve a flux<br />

of 1 Lpm/ft2 (645 LMH).<br />

5. Once the full flux (645 LMH) has<br />

been reached, start the flush.<br />

Millistak+ HC Filters Application Guide<br />

Leak-Testing Installed<br />

Cartridge Assemblies<br />

To check that an assembled stack<br />

has been put together correctly,<br />

use the following procedure:<br />

1. Wet out the filters as described<br />

above.<br />

2. Drain the housing. (This may be<br />

done with a connection to drain<br />

through the inlet to the housing.)<br />

3. Once the housing is empty, apply<br />

a forward pressure of 1 – 2 psid<br />

of air pressure.<br />

4. Wait 3 – 5 minutes for stabilization,<br />

since the air pressure will force<br />

some of the retained water within<br />

the media through the stack.<br />

5. Measure the flow of air into the<br />

housing.<br />

There will be some air flow due to the<br />

low bubble point of the media, and<br />

hence some air permeation through<br />

the media will occur. A typical value<br />

of air flow through a fully-wetted<br />

8-cell stack (1.8 m2 ), with a pressure<br />

differential of 2.0 psid, is 100 to<br />

400 cc/min.<br />