GB<br />
Insert giving set<br />
(See page 11)<br />
Automatic priming using the pump<br />
Keep key<br />
depressed<br />
Program selection<br />
Prime to the end of the giving set<br />
(Release key once priming is complete).<br />
Press key For selection of application programs<br />
see section "Application Programs".<br />
continuous<br />
volume/time<br />
Start<br />
Press key<br />
Stop<br />
bolus<br />
Flashing bars and drops indicate the start of feed/fluid<br />
administration (*).<br />
Press key STOP appears in the display.<br />
Switch off<br />
Keep depressed<br />
for approx.<br />
1 second<br />
In mains mode, the "Plug" symbol is visible in the display.<br />
The pump retains all information when switched off.<br />
This information is saved for 1 month provided the battery was fully<br />
charged at switch off.<br />
(*) In order to detect an upstream occlusion occurred before starting the<br />
pump, the <strong>APPLIX</strong> <strong>Vision</strong> delivers a small initial volume at a higher rate,<br />
right after START, if the door has been previously opened or an upstream<br />
occlusion alarm has been triggered (see Operating conditions p. 35).<br />
4150-2_nu_Applix_<strong>Vision</strong>_GB 19