Alarm Symbol Cause Remedy<br />
Battery alarm in<br />
mains mode<br />
Plug symbol visible in display<br />
28 4150-2_nu_Applix_<strong>Vision</strong>_GB<br />
GB<br />
Batteries are defective. Inform personnel who<br />
service pumps.<br />
Batteries should be changed<br />
by authorised engineers<br />
only!<br />
Plug symbol not visible in display<br />
Mains control indicator on<br />
pump holder does not<br />
come on.<br />
Faulty mains connection.<br />
Mains control indicator on<br />
pump holder comes on.<br />
Pump is not properly<br />
inserted in pump holder.<br />
Contact pins of pump and<br />
pump holder are dirty.<br />
Check that mains cable is<br />
properly fixed in position and<br />
is not damaged in any way.<br />
Insert pump in pump holder.<br />
Ensure it has clicked in to<br />
place.<br />
Remove dirt with cloth and<br />
soapy water or as directed<br />
by hospital policy (see p. 32).<br />
Allow the pump to dry.