Alarm Symbol Cause Remedy<br />

Giving set Giving set inserted<br />

incorrectly or not at all.<br />

Check position of giving set<br />

above and below the pump<br />

mechanism and insert<br />

correctly if necessary.<br />

Wrong giving set inserted. Use recommended <strong>APPLIX</strong><br />

giving set.<br />

Area where clamp is inserted<br />

is contaminated.<br />

Pump door open Pump door not properly<br />

closed at start.<br />

Pump<br />

mechanism<br />

blocked<br />

System fault<br />

"E" plus<br />

numerical code.<br />

Continuous tone<br />

at intervals<br />

Pump door opened after<br />

start.<br />

Pump door removed from its<br />

anchoring.<br />

Remove dirt with cloth and<br />

soapy water or as directed<br />

by hospital policy (see p. 32).<br />

Allow the pump to dry.<br />

Close pump door.<br />

Close pump door.<br />

Re-hang door.<br />

Door mechanism is faulty. Inform personnel who<br />

service pumps.<br />

Faulty pump mechanism. Inform personnel who<br />

service pumps.<br />

Internal equipment fault<br />

(system fault).<br />

Inform personnel who<br />

service pumps.<br />

4150-2_nu_Applix_<strong>Vision</strong>_GB 31