Alarm Symbol Cause Remedy<br />
Giving set Giving set inserted<br />
incorrectly or not at all.<br />
Check position of giving set<br />
above and below the pump<br />
mechanism and insert<br />
correctly if necessary.<br />
Wrong giving set inserted. Use recommended <strong>APPLIX</strong><br />
giving set.<br />
Area where clamp is inserted<br />
is contaminated.<br />
Pump door open Pump door not properly<br />
closed at start.<br />
Pump<br />
mechanism<br />
blocked<br />
System fault<br />
"E" plus<br />
numerical code.<br />
Continuous tone<br />
at intervals<br />
Pump door opened after<br />
start.<br />
Pump door removed from its<br />
anchoring.<br />
Remove dirt with cloth and<br />
soapy water or as directed<br />
by hospital policy (see p. 32).<br />
Allow the pump to dry.<br />
Close pump door.<br />
Close pump door.<br />
Re-hang door.<br />
Door mechanism is faulty. Inform personnel who<br />
service pumps.<br />
Faulty pump mechanism. Inform personnel who<br />
service pumps.<br />
Internal equipment fault<br />
(system fault).<br />
Inform personnel who<br />
service pumps.<br />
4150-2_nu_Applix_<strong>Vision</strong>_GB 31