GB<br />
Nurse call and data interface<br />
The pump holder provides the opportunity of connecting to a nurse call system.<br />
Further information about connection is available from the Technical Service department.<br />
The function is available if the pump is properly engaged in the holder and the holder is<br />
connected to the mains.<br />
Plug the connecting cable for the nurse call function into the connector on the holder and<br />
guide through the lateral cable fixing. Then connect to the nurse call system.<br />
Before using the pump, check that the transfer line for the nurse call system is working<br />
properly by initiating an alarm (e.g. start without giving set) at the pump.<br />
Cable fixing<br />
Mains socket<br />
Nurse call connection<br />
and data interface<br />
4150-2_nu_Applix_<strong>Vision</strong>_GB 9