Alarm Symbol Cause Remedy<br />

Line empty Feed container and giving set<br />

are empty down to the pump.<br />

Giving set not adequately<br />

filled.<br />

Air in sensor area<br />

(with feed container full).<br />

Dirt in sensor area<br />

(lower tube guide).<br />

Giving set not inserted<br />

correctly.<br />

Occlusion Giving set at pump<br />

mechanism is blocked or<br />

kinked.<br />

Upstream Occlusion:<br />

giving set is blocked<br />

between the container and<br />

the pump.<br />

Downstream occlusion:<br />

giving set or feeding tube<br />

is blocked after the pump,<br />

at the patient side.<br />

Continue or end<br />

administration and flush<br />

tube.<br />

Fill giving set to the end.<br />

30 4150-2_nu_Applix_<strong>Vision</strong>_GB<br />

GB<br />

Gently tap bubbles out of the<br />

giving set close to the<br />

sensor. If necessary, fill<br />

giving set again until air<br />

bubbles are removed.<br />

Remove dirt with cloth and<br />

soapy water or as directed<br />

by hospital policy (see p. 32).<br />

Allow the pump to dry.<br />

Check position of giving set<br />

and insert correctly if<br />

necessary.<br />

Open the door, check the set<br />

installation, close the door.<br />

Re-position and verify that<br />

feed flows freely after<br />

adjustment.<br />

Check that the feeding tube<br />

is clear.<br />

Flush tube if necessary.<br />

Check that the feeding tube<br />

is clear.<br />

Flush tube if necessary.