GB<br />

Volume/time feed administration<br />

Volume/time The pump calculates the continuous feed rate from the specified target<br />

volume and the administration time selected. The feed rate is shown in the<br />

display.<br />

Parameter Definition of term Setting range Increments<br />

Administration<br />

time:<br />

Time during which feed is<br />

administered.<br />

Target volume: Total volume of feed selected. An<br />

alarm sounds when the target volume<br />

is reached.<br />

Example settings Results<br />

Target volume Duration of administration Rate<br />

1. 500 ml 7 h 71 ml/h<br />

2. 1500 ml 8 h 188 ml/h<br />

Example 2<br />

200<br />

150<br />

100<br />

50<br />

Rate (ml/h)<br />

1… 24 h 0.5 h<br />

1 2 3 4 5 6 7 8<br />

1… 5000 ml 1 or 10 ml<br />

Time (h)<br />

4150-2_nu_Applix_<strong>Vision</strong>_GB 23