Volume/time feed administration<br />
Volume/time The pump calculates the continuous feed rate from the specified target<br />
volume and the administration time selected. The feed rate is shown in the<br />
display.<br />
Parameter Definition of term Setting range Increments<br />
Administration<br />
time:<br />
Time during which feed is<br />
administered.<br />
Target volume: Total volume of feed selected. An<br />
alarm sounds when the target volume<br />
is reached.<br />
Example settings Results<br />
Target volume Duration of administration Rate<br />
1. 500 ml 7 h 71 ml/h<br />
2. 1500 ml 8 h 188 ml/h<br />
Example 2<br />
200<br />
150<br />
100<br />
50<br />
Rate (ml/h)<br />
1… 24 h 0.5 h<br />
1 2 3 4 5 6 7 8<br />
1… 5000 ml 1 or 10 ml<br />
Time (h)<br />
4150-2_nu_Applix_<strong>Vision</strong>_GB 23