11.3.3 Connection of non-<strong>CS</strong> analogue sensors<br />
Own analogue sensors 0/4…20 mA, 0…1/10V, Pt100, Pt1000 can be connected to the input<br />
channels 3, 4, 5 and 6. For this purpose the specific sensor datea have to be stored in <strong>DS</strong><br />
<strong>300</strong> by means of the <strong>DS</strong> <strong>300</strong> <strong>mobile</strong> <strong>version</strong> configuration software.<br />
Attention: Maximum current supply 12 VDC / 25 mA! If the non-<strong>CS</strong> analogue sensors<br />
have a higher current consumption please use an own power supply!<br />
Connection cables:<br />
<strong>CS</strong> <strong>Instruments</strong> offers connection cables with open ends. Order no. 0553 0110, cable length:<br />
5 m.<br />
Connection scheme:<br />
Attention: <strong>DS</strong> <strong>300</strong> <strong>mobile</strong> <strong>version</strong> is configured on our premises. If you connect your<br />
own analogue sensors <strong>DS</strong> <strong>300</strong> has to be configured newly by means of the<br />
<strong>DS</strong> <strong>300</strong> configuration software.<br />