30.06.2013 Views

GDN Calibration Manual - Kele

GDN Calibration Manual - Kele

GDN Calibration Manual - Kele

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

CALIBRATION PROCEDURE VA301IRF<br />

APRIL 1 ST , 2004<br />

IF YOUR TRANSMITTER WAS PURCHASE BEFORE APRIL 1 ST 2004, CONTACT OUR TECHNICAL SUPPORT LINE AT 1-800-563-2967<br />

CALIBRATION PROCEDURE<br />

VA301IRF<br />

1. WARM-UP PERIOD<br />

Turn the unit on for a minimum of fifteen (15) minutes.<br />

2. CONNECTING THE HARDWARE<br />

Plug the calibration adaptor onto the gas sensor inlet.<br />

Screw the regulator to the calibration span gas cylinder or air cylinder for the zero adjustment.<br />

Connect the regulator outlet to the calibration adaptor with the 1/8” I.D. polymer tubing.<br />

3. ADJUSTING THE ZERO (IF REQUIRED)<br />

This must be achieved in a well ventilated area where there is no presence of refrigerant.<br />

If you suspect the presence of refrigerant in the area, inject air at a flow rate of 100ml/min (air gas bottle<br />

available at Vulcain 103L: #501008 or 17L: #501036), let the gas flow for 1 minute before going on.<br />

Press Enter to get into the menu.<br />

Use the Up or Down button until the password VA is reached.<br />

Press Enter to acknowledge.<br />

Use the Up or Down arrow until the Set Zero field is reached.<br />

Press Enter to acknowledge.<br />

Press Enter again to activate GoCalib.<br />

Wait message appears, the transmitter will go into zero calibration.<br />

When the Menu Set Zero message appears, the zero calibration is complete.<br />

Use the Up or Down arrow until the Quit field is reached.<br />

Press Enter to exit.<br />

4. ADJUSTING THE SPAN<br />

Turn on the regulator.<br />

The calibration span gas is now flowing into the unit.<br />

Let the gas flow for a minimum of 1 minute.<br />

After one minute, the reading should be stabilized.<br />

Press Enter to get into the menu.<br />

Use the Up or Down button until the password VA is reached.<br />

Press Enter to acknowledge.<br />

Use the Up or Down arrow until the Set Span field is reached.<br />

You can now set the Span gas value (in PPM) using the Up or Down button.<br />

Press Enter to acknowledge.<br />

Press Enter again to activate GoCalib.<br />

Wait message appears, the transmitter will go into span calibration.<br />

When the Menu Set Span message appears, the calibration is complete.<br />

Use the Up or Down arrow until the Quit field is reached.<br />

Press Enter to exit.<br />

PAGE 1 OF 3 © COPYRIGHT VULCAIN ALARME INC. DI0141

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!