charge-up, we modified the structure of the holes and the surface of the electrode.<br />
Table 2-8 shows a summary of such modifications together with observed<br />
pulse-height shifts under various injection rates of particles with the energy of about 1<br />
MeV/nucleon. THGEM #1 indicates the THGEM previously developed in Ref. 2-25,<br />
which has an insulator rim of 100 µm in width around a hole. THGEM #2 has no rim<br />
and the gain shift was suppressed to be less than 12 % under the 4 He injection rate of<br />
10 4 pps. However, a discharge often occurred in several hours from the beginning of the<br />
injection, and the gain became unstable once the discharge occurred. In order to avoid<br />
the discharge, the diameter of the hole of THGEM#3 was reduced to be smaller than of<br />
THGEM#2.<br />
Table 2-8. Summary of modified THGEMs together with pulse height shifts at various injection<br />
rates. THGEM#1 indicates the previously developed THGEM [2-25]. THGEM#2 has no rim on the<br />
hole of 500 µm in diameter. The diameter of the hole of THGEM#3 was smaller than for<br />
THGEM#2. The surface of copper electrodes was coated with gold for THGEM#4. ‘n/a’ indicates<br />
not-measured.<br />
Pulse<br />
height<br />
THGEM# #1 #2 #3 #4<br />
Hole diameter [µm] 300 500 300 300<br />
Rim yes no no no<br />
Hole pitch [µm] 700 700 700 700<br />
Electrode surface Au Cu Cu Au<br />
4 He (E = 5.4 MeV, 10 0 pps ) 100 a few n/a n/a<br />
4 He (E = 5.4MeV, 10 4 pps ) n/a 12 1~5 n/a<br />
shift (%) 12 C (E = 13.2 MeV, 10 5 pps) n/a n/a 8 < 3<br />
Fig. 2-43. Pulse-height shifts under the 12 C-beam injection rate from 400 pps to 10 5 pps using<br />
THGEM #3 (left panel) and THGEM #4 (right panel). The horizontal axis indicates time from the<br />
beginning of the beam injection.<br />
49