accuracy. The results showed that the Iep ES measurements were within 2.4 % <strong>of</strong>the<br />

expected phosphorus concentrations, and 7 % <strong>of</strong>the expected sodium concentrations.<br />

Each original sample was diluted to give two aliquots within the range 50-200 ppm in Na<br />

and P, and agreement was within ±3 %.<br />

The results that were received from ICP-ES analysis were the elemental<br />

concentrations (ppm) <strong>of</strong>the diluted samples. Using a Micros<strong>of</strong>t Excel spreadsheet and<br />

the mass dilution factors <strong>of</strong>each diluted sample, the sodium and phosphorus molality<br />

concentrations <strong>of</strong>each original sample taken from the stirred reactor were calculated.<br />

2.3.4 Electron Microscopy<br />

From the synthesis experiments, the solid reaction products <strong>of</strong>specific interest to<br />

this study were analyzed by scanning electron microscopy (SEM) at an accelerating<br />

voltage <strong>of</strong>20 kV in a Hitachi S570 SEM equipped with a Tracor Northern 5500 cnergy­<br />

dispersive X-ray analyzer and a Microtrace Model 70152 silicon X-ray spectrometer,<br />

located in the Department <strong>of</strong> BiologylMUN. Detector/sample positioning gave an<br />

effective take<strong>of</strong>fangle <strong>of</strong>30" and SEM images were recorded on Polariod Type 655<br />

PositiveJNegative film. Samples were placed on double sided sticky tabs on aluminium<br />

stubs and carbon coated in a Denton 502A High Vacuum Evaporator before SEM<br />

analysis.