7.3 The Source Runs<br />
Source no. 1 was mounted with a short nylon cord directly to the steel rod. With this source<br />
we could measure only positions on the z-axis. With source no. 2 we used a short coupling to<br />
a hinge <strong>and</strong> a 48 cm lever arm that could be lifted up to a horizontal position using a nylon<br />
pull-string (see fig. 7.4). The source was mounted with a short cord at the end <strong>of</strong> the lever arm.<br />
With this source we measured points on the z-axis as well as points with a horizontal distance<br />
<strong>of</strong> 48 cm from the z-axis.<br />
Figure 7.4: Stainless steel rod with hinge for the positioning <strong>of</strong> the sources inside the Inner<br />
Vessel. If the nylon string is loose, the lever arm hangs vertically down. If the nylon string is<br />
pulled upwards, the lever arm is brought into a horizontal position.<br />
The nominal source positions have a rather big uncertainty:<br />
- As the source was mounted on a steel rod in principle the position on the z-axis should<br />
be known very accurately. The measurement was always done relative to the top flange<br />
101