7.3 The Source Runs<br />
layers <strong>of</strong> plastic bags <strong>and</strong> black felt to prevent light from entering. It cannot be excluded<br />
that the direction was turned a little bit during this procedure.<br />
- When pulling up the lever arm to measure <strong>of</strong>f-axis-points, the vertical part <strong>of</strong> the steel<br />
rod was no longer perfectly vertical, but slightly tilted. This introduces an additional<br />
uncertainty in the position.<br />
Due to these effects, the relative error in the nominal position is estimated to be<br />
¡Ü ¡Ý ¦ Ñ for <strong>of</strong>f-axis-points<br />
¡Ü ¡Ý ¦ Ñ for points on z-axis<br />
¡Þ ¦ Ñ<br />
Nominal source positions<br />
Run x (cm) y (cm) z (cm) rÜÝ (cm) rÜÝÞ (cm)<br />
source 1 791 0 0 +32 0 32<br />
793 0 0 0 0 0<br />
795 0 0 +100 0 100<br />
source 2 797 0 +48 0 48 48<br />
798 -48 0 0 48 48<br />
799 +27 -40 0 48 48<br />
800 +40 -27 -30 48 57<br />
801 -40 +27 -30 48 57<br />
802 0 0 -78 0 78<br />
803 0 0 -83 0 83<br />
804 0 0 -75 0 75<br />
805 0 0 -62 0 62<br />
807 -40 +27 -56 48 74<br />
809 -40 -27 -56 48 74<br />
810 27 -40 +11 48 49<br />
Table 7.1: The source positions during the different runs (the runs not mentioned here were only<br />
very short runs to verify the source position). The coordinates x, y, z are given in cm, relative to<br />
the ‘nominal center’ <strong>of</strong> the IV.<br />
103