Connection D - Controller-to-Pump Connection<br />
To connect the <strong>controller</strong> to the pump, an appropriate<br />
interconnection cable must be prepared to connect the<br />
P10 connector of the <strong>controller</strong> with the pump<br />
connector.<br />
The following figure shows the pump output connector<br />
configuration where pins:<br />
− A-F = pump temperature sensor<br />
− B-C-D = 80 Vac 3-phase output to pump motor<br />
stator<br />
− E = ground.<br />
P10<br />
PHASE R<br />
PHASE S<br />
PHASE T<br />
PUMP TEMP. SENSOR<br />
PUMP TEMP. SENSOR<br />
GND<br />
TECHNICAL INFORMATION<br />
A<br />
F B<br />
E<br />
D<br />
C<br />
CONTROLLER SCREEN<br />
PUMP<br />
Controller to pump cable<br />
TURBO-V<br />
PUMP<br />
CONNECTOR<br />