Connection D - Controller-to-Pump Connection<br />

To connect the <strong>controller</strong> to the pump, an appropriate<br />

interconnection cable must be prepared to connect the<br />

P10 connector of the <strong>controller</strong> with the pump<br />

connector.<br />

The following figure shows the pump output connector<br />

configuration where pins:<br />

− A-F = pump temperature sensor<br />

− B-C-D = 80 Vac 3-phase output to pump motor<br />

stator<br />

− E = ground.<br />

P10<br />

PHASE R<br />

PHASE S<br />

PHASE T<br />

PUMP TEMP. SENSOR<br />

PUMP TEMP. SENSOR<br />

GND<br />

TECHNICAL INFORMATION<br />

A<br />

F B<br />

E<br />

D<br />

C<br />

CONTROLLER SCREEN<br />

PUMP<br />

Controller to pump cable<br />

TURBO-V<br />

PUMP<br />

CONNECTOR<br />

75 87-900-884-01 (D)