Turbo-V 300HT PCB controller
Turbo-V 300HT PCB controller
Turbo-V 300HT PCB controller
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
USE<br />
General<br />
Make all vacuum manifold and electrical connections and<br />
refer to <strong>Turbo</strong>-V pump instruction manual before to<br />
operating the <strong>Turbo</strong>-V <strong>controller</strong>.<br />
!<br />
To avoid injury to personnel and damage to the equipment,<br />
if the pump is laying on a table make sure it is steady.<br />
Never operate the <strong>Turbo</strong>-V pump if the pump inlet is not<br />
connected to the system or blanked off.<br />
NOTE<br />
The input signal P1 connector should be left in position<br />
including the shipping links if no external connections are<br />
made. The forepump and the <strong>Turbo</strong>-V pump can be<br />
switched on at the same time.<br />
NOTE<br />
When the <strong>Turbo</strong>-V <strong>300HT</strong> pump is baked by a membrane<br />
pump, the Soft Start mode should be deselected.<br />
NOTE<br />
WARNING!<br />
The following instructions are applicable to <strong>Turbo</strong>-V<br />
<strong>300HT</strong> <strong>PCB</strong> Controller only when the optional "hand<br />
held terminal" is installed.<br />
Startup<br />
• Apply the mains to the <strong>controller</strong>.<br />
• The display lights, and shows:<br />
TECHNICAL INFORMATION<br />
R E A D Y F O R L O C A L<br />
S O F T S T A R T<br />
The <strong>controller</strong> with the Soft Start mode allows the<br />
pump to ramp-up to Normal speed slowly with a<br />
minimum ramp-up time of 30 minutes and a<br />
maximum of about 55 minutes.<br />
The Soft Start mode is always operative as default<br />
mode. If it is necessary to deselect this mode refer<br />
to the following paragraph.<br />
If the Soft Start mode is deselected, the ramp-up<br />
will be done within 300 seconds and the display<br />
changes as follows:<br />
P U M P R E A D Y : P U S H<br />
S T A R T B U T T O N<br />
• Press the CURRENT pushbutton and the display<br />
shows (e.g.):<br />
P = X X X W X X ° C<br />
Y Y X X K R P M<br />
where:<br />
• P = is the DC power drawn by the pump (range 0 to<br />
999 Watt)<br />
• KRPM = is the theoretical rotational speed of the<br />
pump as a function of the <strong>controller</strong> output<br />
frequency (range 3 to 56 KRPM)<br />
• °C = is the temperature of the outer ring of the<br />
upper bearing (range 00 to 99°C)<br />
• Y = during operation a selected set point condition<br />
(1 or 2 contrast inverted) appears when the<br />
programmed threshold speed value is not reached.<br />
83 87-900-884-01 (D)